11
7-
Main Power Down:
Set Main Voltages to Off, this routine disables the mechanical switch. To enable the mechanical switch the power to
the extender board has to be turned off and then on.
Use the "Set Address" routine to set address
DUT 1 = 0x64
DUT2 = 0xe4
use write routine to write
0x1
8-
+5V Aux Up
Be sure to have the Global register to on before turning any voltages on or off
Use the "Set Address" routine to set address
DUT 1 = 0x65
DUT2 = 0xe5
use write routine to write
0x0
9-
+5V Aux Down:
Use the "Set Address" routine to set address
DUT 1 = 0x65
DUT2 = 0xe5
use write routine to write
0x1
10-
+3.3V Aux Up:
Be sure to have the Global register to on before turning any voltages on or off
Use the "Set Address" routine to set address
DUT 1 = 0x66
DUT2 = 0xe6
use write routine to write
0x0
11-
+3.3V Aux Down:
Use the "Set Address" routine to set address
DUT 1 = 0x66
DUT2 = 0xe6
use write routine to write
0x1
12-
RST# setup:
Data[20..0] provides a delay time up to 5.368709 seconds in 2.56 microseconds increment.
D23 = 0, sets the reset time to default of about 150 ms
D23 = 1, sets the reset time to the value defined by D0 to D20
Use the "Set Address" routine to set address
DUT 1 = 0x71
DUT2 = 0xf1
use write routine to write D[7..0]
Use the "Set Address" routine to set address
DUT 1 = 0x72
DUT2 = 0xf2
use write routine to write D[15..8]
Use the "Set Address" routine to set addres s
DUT 1 = 0x73
DUT2 = 0xf3
use write routine to write D[23, 20..16]