OPERATOR’S MANUAL
HB3343-01
PRINTED IN ENGLAND
Страница 1: ...TUFF PERSONAL AIR SAMPLING PUMPS Go to Contact Details OPERATOR S MANUAL HB3343 01 PRINTED IN ENGLAND...
Страница 2: ...53 E mail info casellaUSA com Web www casellaUSA com CASELLA ESPANA S A Pol gono Eur polis Calle C n 4B 28230 Las Rozas Madrid Spain Phone 34 91 640 75 19 Fax 34 91 636 01 96 E mail online casella es...
Страница 3: ...1 Introduction 5 2 Charging the Tuff Unit Batteries 6 3 Operation Symbols 7 4 Operation 9 5 Configuration 14 6 Calibration 17 7 Renew Inlet Adapter Filter 19 8 Servicing 20 9 Fault Finding 21 10 Pump...
Страница 4: ...ed the instrument must be returned immediately to Casella CEL or to a Casella CEL Approved Agency for repair The warranty does not extend to cleaning or general servicing of the instrument Pumps which...
Страница 5: ...the U K Health and Safety Executive s publication MDHS14 and in other international reference methods The information contained in this handbook relates only to the operation of Casella CEL sampling...
Страница 6: ...t Battery to Pump and then to the Charger Insert pump into charging cradle as shown above Fast charge starts and the red LED on the pump flashes When fully charged the blue LED will illuminate The pum...
Страница 7: ...ee times within 3 seconds to enter Partial Lock Mode In this mode the pump can only be started and stopped Fully Locked Mode This can only be activated in Run Mode Press the ON OFF key three times wit...
Страница 8: ...p is running the Sampler will scroll through the Run Mode Menu options and the blue led will be flashing Current run data will be displayed To stop the pump press and hold the ENTER key on any screen...
Страница 9: ...DOWN keys to set the total sampling duration e g 8 hours Press ENTER to continue to the Exposure time setting Use the UP DOWN keys to set the Exposure Time e g 1 hour The settings entered will provide...
Страница 10: ...ode Menu Options See Fig 14 for details until PRG1 screen is displayed Press ENTER to continue to the ON 1 screen Use the UP DOWN keys to review the program set parameters ON1 is 08 00 OFF1 is 12 00 P...
Страница 11: ...o maximum of 3l min whilst Tuff 4 models will adjust up to 4 5L min flow rates Press ENTER to continue Pump should then be connected to external flow calibration device For External Calibration detail...
Страница 12: ...e TUFF inlet connect the following constant pressure regulator low flow adaptor sorbent tube holder and a flow meter such as the Dry Flow or Dry Cal shown in Figure 14 4 Break both ends off a sorbent...
Страница 13: ...nd hold the UP arrow This allows access to the configuration screens After approximately 8 seconds the Set Language screen is displayed Figure 15 Access Configuration Screens B SET LANGUAGE Fig 16 Use...
Страница 14: ...e 21 Service Life and Serial Number Screen H CALIBRATE YES NO Default Calibration Tuff Plus and Pro Only Fig 22 Use UP DOWN keys to SELECT Yes No Activates Calibration Mode which allows the entire cal...
Страница 15: ...R to save the new setting and continue Figure 24 Low Set Adjust Screen K ADJUSTMENT OF HIGH SET POINT Fig 25 Use UP DOWN keys to alter the high set point in l min Press ENTER to continue Figure 25 Hig...
Страница 16: ...on Attach a sampling head and flow tube or other flow measuring device to the pump s inlet nozzle to measure the actual flow as shown in Figures 28 29 and 30 Press the UP DOWN keys to adjust the flow...
Страница 17: ...F Hose Nominal Diameter 5mm G Hose Nominal Diameter 5mm H TUFFTM Pump Figure 29 Calibration with Digital Calibrator A Cyclone Sampling Head B TUFFTM Pump C Hose Nominal Diameter 5mm D Typical Digital...
Страница 18: ...ironment can reduce this time considerably 1 Remove the Inlet nozzle 2 Discard the Filter Element 3 Fit a new Filter Element 4 Fit and hand tighten the Inlet nozzle Figure 31 Renewing Inlet Filter Pac...
Страница 19: ...to repair your instrumentation Should you wish factory repair assistance send your equipment in a package equivalent to the original packaging Insure to full value and ship prepaid Include a letter g...
Страница 20: ...ump runs fast Control error Damaged flow pressure sensor return for repair Water ingress clean or return for repair Tubes to sensor squashed or damaged return for repair Electrical fault return for re...
Страница 21: ...n be left running in the background on the Windows system as an icon When a pump comes within range of the infrared computer link the application will wake up and download data from the pump and store...
Страница 22: ...he on screen instructions Once installation is complete you will find the Pump Manager icon on your Windows desktop UN INSTALL 1 Click the Start button on the task bar 2 Select the Settings option and...
Страница 23: ...to the right of the Status Bar at the bottom of the PC display This indicates that the device is waiting to communicate with an TUFF Pump Best communication with the pump is achieved when the IR link...
Страница 24: ...the display to show that the software is active in background Run on Startup Enable this option to start Pump Manager automatically whenever Windows is started Exit Standard dialog All data will be id...
Страница 25: ...y their details on the New Person dialog Delete Standard dialog Properties Offers the following options Name Alphanumeric characters of the person s name can be entered ID Alphanumeric characters to i...
Страница 26: ...are continues transmitting the request and listening for a reply However when a pump containing data is within range it will be detected and the download process started automatically A Data Download...
Страница 27: ...ry field a New Pump message will be displayed Fig 35 and the new pump added automatically to the application The operator should add the remaining pump properties as soon as possible A TUFF Standard m...
Страница 28: ...in the Data area with headings that give a summary of information from the event 4 To inspect further data from any event double click on the sample 5 An Event Properties dialog is displayed that con...
Страница 29: ...allows calibration information to be added and applied to the data Use calibration data Tick to use the displayed information Calibrator S N Allows a serial number to be added Initial flow Field for a...
Страница 30: ...e identified Date The date of the sample or of the report can be entered Audited by The data report checker can be identified Date The date of checking can be entered Figure 41 Report Page Layout 9 Us...
Страница 31: ...be edited then sent to the pump at the same time as follows 1 Select the Pump menu 2 Select the Program option The Pump Properties dialog will be displayed showing program types Duration TWA Prog 1 a...
Страница 32: ...r the run A day of the week can be specified where All days allows the instrument to be used on any day Start and end times can be specified in 1 minute steps up to 24 hours and 59 minutes 11 Enter a...
Страница 33: ...pump found to program ERROR Cannot set pump s time ERROR Cannot set pump s date ERROR Could not send program to the pump ERROR Could not get the pump s serial number ERROR The pump has changed during...
Страница 34: ...ice Interval Typically 2500 hrs Operating Temperature 5o C to 45o C Storage Temperature 10o to 50o C Charging Technique Drop in Intelligent fast charger employing dT dt termination with safety time ou...
Страница 35: ...7 17 30 114 21 13 40 138 18 11 50 159 15 10 60 181 13 8 70 195 12 8 2 L min 2 58 42 26 5 64 38 24 10 78 31 20 20 103 24 15 30 128 19 12 40 148 16 10 50 168 14 9 60 184 13 8 70 197 12 8 2 5 L min 2 75...
Страница 36: ...E 25mm 1 2um MCE 37mm GFA Tuff Pump 1 7Ah Battery Performance vs Flow Rate 0 10 20 30 40 50 60 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 Back pressure cm H2O Run time h 1 L min 2 L min 3 L m...
Страница 37: ...nal Sampling Pumps are designed to comply with the EMC Directive 89 336 EEC of the European Union They have been tested according to the standard delivery schedule and comply with the following standa...
Страница 38: ...rating procedures may differ slightly from the operating procedures given in this manual If there are any questions you are encouraged to contact Casella CEL for clarification While every care is take...