21 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, check that
each of the following requirements is satisfied:
Power source:
The wall power outlet must meet the power requirements listed on the unit data
plate.
Model
Voltage
Amps
Frequency
CVO-5
110 – 120
14
50/60 Hz
CVO-5-2
220 – 240
7
50/60 Hz
CVO-10
220 – 240
10
50/60 Hz
•
Wall power sources must be
protective earth grounded
and
single phase
.
•
The source must conform to all national and local electrical codes.
•
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to the
oven may result if supplied voltage varies more than 10%.
•
The recommended wall circuit breaker for these units are:
o
CVO-5:
15 amps
o
CVO-5-2, CVO-10:
15 amps
•
A separate circuit for the oven is recommended to prevent possible loss of product due to
overloading or failure of other equipment on the same circuit.
Power Cord:
The unit must be positioned so that all end-users can quickly unplug the oven in the
event of an emergency. Always use the same type of power cord that was shipped with the oven.
Oven Fuses:
All fuses must be present and not blown for the oven to operate. The cause of a blown
fuse should always be determined and fixed prior to returning the oven to operation.
CVOs ship with the following 5x20mm fuses installed:
Model
Type
Quantity Location
CVO-5
T16A 250V
1
Power cord inlet
CVO-5-2
T10A 250V
2
Power cord inlet and fuse holder adjacent to the cord inlet.
CVO-10
T10A 250V
2
Power cord inlet and fuse holder adjacent to the cord inlet.
Continued on next page
CVO-5
CVO-5-2
CVO-10-2
Содержание CVO-10
Страница 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Страница 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Страница 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Страница 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Страница 56: ......