CAS CI-2001A Скачать руководство пользователя страница 1

 

2

 

CI-2001AB  국문

Содержание CI-2001A

Страница 1: ...2 CI 2001AB 국문 ...

Страница 2: ...AUTIONS 6 FEATURES 6 TECHNICALSPECIFICATION 7 DIMENSIONS 9 FRONT PANEL CI 2001A 10 FRONT PANEL CI 2001B 13 REAR PANEL 16 INSTALLATION 17 TEST MODE 18 CALIBRATION MODE 21 SET MODE 25 SERIALINTERFACE 30 ERROR MESSAGE 34 ...

Страница 3: ... might cause failure or electricshockuponelectricleak Do not damage process excessively jerk bendortwistthepowercord It might damage the power cord to causefireorelectricshock Keep any combustible spray or fire sourceaway Itmightcausefire Do not spraywater to the outside of the productoruseitinanyhumidplace It might deteriorate the insulation of electric parts that can cause the electric shock fir...

Страница 4: ...voidanysuddenshocktotheproduct It might damage the product to fail the accurateweighing Find a proper place to attach the rubber pad at the bottom of the indicator whichwasshippedtogether Do not use the product at a place with sudden temperature changes or severe vibrations It might cause the weighing error or failure Donotinstalltheproduceataplacewiththeexcessiveelectromagneticwave Itmightcauseth...

Страница 5: ...play functions to help users understand the product Please use the product right and sufficiently utilize functions of CI 2000 series as you read this manual thoroughly before using CI 2000 series FEATURES 1 Features Appropriate for weight and measurement system Easy operation and various options Simple Full Digital Calibration SPACTM Single pass automatic span calibration WATCHDOG circuitry Syste...

Страница 6: ...o 4 x 350Ω load cells Zero adjust range 0 05 mV 5 mV 2 μV D NTEP OIML KS Input Sensitivity 0 5 μV D Non NTEP OIML KS System Linearity 0 01 F S A D internal resolution 1 200 000 A D external resolution 1 5 000 NTEP OIML KS 1 30 000 Non NTEP OIML A D conversion speed 10times sec ...

Страница 7: ...ACAdapter minus sign CI 2001A Zero Tare Gross Net Stable lb kg Power Consumption Operating Temperature CI 2001B Zero Net Stable lb kg Overall Dimensions AC 110V 220V DC 6V 12V 850mA CI 2001A 10 W Weight Span calibration CI 2001B 1 W Display 10 to 40 Size of letter 85mm x 186mm x 58mm Display below zero 0 5 kg Option Standard Serial Interface RS 232 Option 1 Serial Interface RS 422 485 Option 2 Pan...

Страница 8: ...9 DIMENSIONS WALL MOUNT TYPE CI 2001A B STANDARD PANNEL MOUNT TYPE CI 2001A B P Cutting Size 166mm x 76mm ...

Страница 9: ...lb Lamp ON when the weight unit is pound lb kg Lamp ON when the weight unit is kilogram kg STABLE Lamp ON when the weight is stable TARE Lamp ON when the tare weight is stored GROSS Lamp ON when the current weight is GROSS weight NET Lamp ON when the current weight is NET weight ZERO Lamp ON when the current weight is 0 kg 0 lb ...

Страница 10: ...ight is memorized as tare weight If you press TARE key in unload condition Tare setting is released GROSS NET KEY Display gross and net weight by turn GROSS lamp on gross weight NET lamp on net weight In case tare weight is REGISTERED tare and item s total weight is G weight and only item s weight is N weight Kg lb KEY Toggles between lb and kg units PRINT KEY PRINT Key ENTER KEY Total Print KEY I...

Страница 11: ...ssing the PRINT key and SET mode starts 4 How to enter SET mode Turn on the Power while pressing the ENTER key and SET mode starts 5 How to enter CAL mode Turn on the power while pressing the CALswitch on the rear panel of the indicator and CAL mode starts ...

Страница 12: ...LIGHT TARE NET GROSS lb kg PRINT CAPACITY 1 Display LAMP Lb ON when the weight unit is pound lb kg ON when the weight unit is kilogram kg ON when the weight is stable ZERO ON when the current weight is 0 kg 0 lb NET ON when the current weight is NET weight ...

Страница 13: ...morized as tare weight If you press TARE key in unload condition Tare setting is released GROSS NET KEY Display gross and net weight by turn GROSS lamp on gross weight NET lamp on net weight In case tare weight is REGISTERED tare and item s total weight is G weight and only item s weight is N weight KEY Total Print KEY In case of F09 1 Hold KEY In case of F09 2 PRINT KEY PRINT Key Total Print key ...

Страница 14: ...ssing the PRINT key and SET mode starts 4 How to enter SET mode Turn on the Power while pressing the LIGHT key and SET mode starts 5 How to enter CAL mode Turn on the power while pressing the CAL switch on the rear panel of the indicator and CAL mode starts ...

Страница 15: ...annel Description RS 232C PORT Serial interface port Computer printer LOADCELL Port for connecting load cell POWER Port for DC power DC 12V adaptor are available CALS W Used in calibration starts ON OFF Power ON OFF switch ...

Страница 16: ...mended resolution 2 mV 1 1 000 Max 4 mV 1 2 000 Max 10 mV 1 5 000 Max Manufacturer s wire colors Connector Corperation No 1 EX No 2 EX No 3 SIG No 4 SIG No 5 Shield CAS RED WHITE GREEN BLUE SHIELD KYOWA RED BLACK GREEN WHITE SHIELD INTERFACE RED BLACK GREEN WHITE SHIELD P T RED BLACK GREEN WHITE SHIELD BLS GREEN BLACK WHITE RED YELLOW SHOWA RED BLUE WHITE BLACK SHIELD SHINKOH RED BLACK GREEN WHITE...

Страница 17: ... the left by 1 place of the set value Initialize Setting value ENTER Move into next menu CI 2001A LIGHT Move into next menu CI 2001B 3 TEST MENU TEST 1 TEST 5 TEST 1 KEYTEST TEST 2 DISPLAYTEST TEST 3 Loadcell test andA D conversion TEST TEST 4 Serial interface test RS 232 TEST 5 Printer test TEST 1 FUNCTION KEYTEST KEY DISPLAY DESCRIPTION ENTER LIGHT Next Menu Other Keys Perform TEST TEST 1 condit...

Страница 18: ...y REF 1 Program is automatically shifted to menu selection mode after completing TEST 3 FUNCTION A D Converer TEST L C TEST KEY DISPLAY DESCRIPTION ENTER LIGHT Next Menu Other Keys Perform TEST TEST 3 condition Dispaly digital value of current weight This value means converted digital value REF 1 Check whether digital value is changing If the digital value is fixed or zero is displayed please chec...

Страница 19: ...rd and check if indicator receives no 1 Send no 1 in indicator keyboard and check if computer receives no 1 REF 3 Do this test after baud rate is specified in SET mode F11 TEST 5 FUNCTION PRINTTEST KEY DISPLAY DESCRIPTION ENTER LIGHT Next Menu Other Keys Perform TEST TEST 5 condition No error in printer Check printer connector REF 1 Please confirm that F02 SET MODE value is set as 1 before a test ...

Страница 20: ...Increase the first place set value to 1 Move to the left by 1 place of the set value PRINT kg lb version kg only version Initial 0 of the set value ENTER Move into next menu CI 2001A LIGHT Move into next menu CI 2001B 3 CALIBRATION MENU CAL 1 CAL 5 CAL1 Maximum Capacity CAL2 Minimum Division CAL3 Setting Weight CAL4 Zero Calibration CAL5 Span Calibration ...

Страница 21: ...maximum capacity means the maximum weight that scale can measure CAL 2 FUNCTION SET Minimum Division Range 0 0005 100 kg lb KEY DISPLAY DESCRIPTION Input the next division ENTER LIGHT Store and move into next menu CAL2 condition 1 kg 0 01 kg REF 1 The minimum division means the value of one division REF 2 External resolution is obtained by division the min division by the maximum capacity Set the ...

Страница 22: ...ity Error message ERR 22 will occur REF 3 If the Setting Weight over the Maximum Capacity Error message ERR 23 will occur CAL 4 FUNCTION Zero Calibration KEY DISPLAY DESCRIPTION ENTER LIGHT Zero calibration and Move into next menu CAL4 condition Unload the tray and press ENTER Under zero calibration Zero calibration is completed REF 1 If Zero calibration is done without any error GOOD message is d...

Страница 23: ...bration Span calibration is completed Press ENTER LIGHT key Save Exit CALMode REF 1 If Span calibration is done without any error GOOD message is displayed The weight of setting weight is displayed on LCD screen Check the weight REF 2 If the span is low Error message ERR 24 is displayed Calibrate with lower resolution 4 SEALING METHOD Install the seal on the wire loop as shown in below Figure 4 ...

Страница 24: ...ALUE CONVERSION MENU F01 F19 F01 Zero Key Operating Range Only CI 2001B F02 Serial Port RS232 Usage F03 Automatic zero tracking compensation F04 Digital Filter F07 Weight Back up Power onActual Weight F08 PRINT Key Usage Only CI 2001A F09 LIGHT Key Usage F10 Device ID F11 Designation of Serial Interface Baud rate F12 Designation of Serial Interface output mode F13 Set HOLD type F14 Select of clock...

Страница 25: ...onnection with Serial printer Automatic Zero Tracking 0 NoneAutomatic zero F03 1 9 1 0 5 Digit 9 4 5 Digit Autozero tracking will automa tically bring the display back to 0 when there are small d eviations Digital Filter F04 1 9 1 Less Vibration 9 MuchVibration Adjust the set value according to the condition Weight Backup 0 Weight back up is off Power on Zero F07 1 Weight back up is on PRINT Key U...

Страница 26: ... 2 HOLD Key Key Usage Only CI 2001B 0 No Use 1 Total Print Key F09 2 HOLD Key Device ID F10 00 99 00 Device ID 0 99 Device ID 99 It is used the ID of indicator when system is connected Baud Rate 0 600 bps 1 1200 bps 2 2400 bps 3 4800 bps 4 9600 bps F11 5 19200 bps ...

Страница 27: ... Set Hold type 0 Average hold Compute the average weight of oscillating weights 1 Peak hold Compute the maximum weight among oscillating weights F13 2 Sampling Hold Select Option Clock only CI 2001B 0 Not use Clock F14 1 use Clock Change Date Time Ex 2014 07 26 13 10 01 Display Meaning Ex C1 14 Year 14 C2 07 Month 07 C3 26 Day 26 C4 13 Hour 13 C5 10 Minute 10 Increase of no Shift of digit ENTER LI...

Страница 28: ...Feed F17 0 9 0 No Line Feed 9 9 Line Feed Set Line Feed Set Model Select Only CI 2001B 0 Set Model CI 2001B 1 Set Model CPS F18 2 Set Model WCS SetAuto Print Only CI 2001B 0 No UseAuto Print F19 1 UseAuto Print ...

Страница 29: ...of CI 20001 A B Serial port of the computer RXD 3 O O 3 Transmit Data TXD 2 O O 2 Receive Data GND 7 O O 7 Signal Ground O 8 Carrier Detect O 20 Data Terminal Ready O 6 Data Set Ready O 4 Request to Send O 5 Clear to Send 9 pin port male 25 pin port female RS 232C port of CI 20001 A B Serial port of the computer Sub DisplayConnection RXD 3 O O 2 Transmit Data TXD 2 O O 3 Receive Data GND 7 O O 7 S...

Страница 30: ...ver can receive data selecti vely which indicator send Device ID is selected in F22 Weight Data 8 byte 1 13 5 kg 1 3 5 2 135 kg 1 3 5 3 135 kg 1 3 5 EachASCII code of Weight transmitted by 8 byte Ex 0x20 Simple Interface Program GWBASIC Language 10 OPEN COM1 9600 N 8 1 As 1 20 IF LOC 1 0 THEN 60 30A INPUT 1 1 40 PRINTA 50 GOTO 20 60 B INKEY IF B THEN 20 70 PRINT B 80 PRINT 1 B 90 GOTO 20 ...

Страница 31: ... COM1 cprintf BIOSCOM ESC to exit n while DONE status bioscom 3 0 COM1 if status DATA_READY if out bioscom 2 0 COM1 0x7F 0 putch out if kbhit if in getch x1B DONE TRUE bioscom 1 in COM1 return 0 CP 7000SeriesPrinterConnection RXD 3 O O 3 Transmit Data TXD 2 O O 2 Receive Data GND 7 O O 14 Signal Ground 9 pin port male 15 pin port female RS 232C port of CI 20001 A B CP 7000 Series Printer Port ...

Страница 32: ...O O 5 Signal Ground 9 pin port male 9 pin port female RS 232C port of CI 20001 A B ND Series Serial Printer Port Option RS 422 485 Serial Interface Transmit mode Same as RS 232C interface COM1 Data Format Same as RS 232C interface COM1 ConnectingmethodofRS 422Port ConnectingmethodofRS 485Port ...

Страница 33: ...inter connection Trouble shooting Check with printer connector If there is no problem with printer and printer connector please request A S to head office Err 08 Reason The ZERO key or TARE key is adjusted not to be operated under the unstable condition Trouble shooting Press Zero or Tare key in stable condition Err 09 Reason Current weight deviates from zero range Trouble shooting Press the ZERO ...

Страница 34: ...the calibration menu and reset the calibration using weight Err 13 Reason The zero range deviates from the set range Trouble shooting Confirm that there is nothing on the weighing platform If nothing exist do calibration in CALmode Over Reason The weight on platform is too heavy to be measured Trouble shooting Do not load the item exceeds the maximum tolerance If the load cell is damaged the load ...

Страница 35: ...solution should be below 1 10 000 Err 22 Reason The weight for span calibration is set to be lower than 10 of the maximum capacity of the scale Trouble shooting Set the weight for span calibration in CAL3 to be more than 10 of the maximum capacity Err 23 Reason The weight for span calibration is set to be exceeded 100 of the maximum capacity of the scale Trouble shooting Set the weight for span ca...

Страница 36: ...Voltage Recommended Resolution 2 mV 1 1 000 4 mV 1 2 000 10 mV 1 5 000 Err 25 Reason The load cell output is too large at SPAN calibration Trouble shooting Setting of current resolution is not possible due to the error in load cell Proceed calibration again with less resolution Err 26 Reason The load cell output is too large at ZERO calibration Trouble shooting Check whether the platform empty Pro...