17

SENSITIVITY CALIBRATION 

 

The electronic balance measures mass by electronicity compensating for terrestrial gravitation. 
Since gravitation varies slightly in different regions, span calibration (sensitivity calibration) is required 
when the balance is installed. Temperature also effects balance accuracy, and calibration must be 
performed whenever a significant change occurs. It is good practice to calibrate the balance whenever 
the balance is moved or unexpected shock is applied to the balance such that an article drops on the pan.

 

 

1. Setting the value of sensitivity calibration weight 

 

In this balance, the value for sensitivity calibration weight can be set freely within the specified range. 
Using the weight having known value, sensitivity calibration can be made.   
Set the weight value using at sensitivity calibration as follows.

 

 

Step 

 

(1) Following the menu selection, press the 

CAL/MENU

 key to 

make the display “ 

 ”. 

(2) Press the 

TARE 

key to set the balance to weight set mode.

 

(3) The weight value currently set is displayed and the digit to be set 

blinks. When the weight value is not changed, press the 

TARE

 key. 

(4) Pressing the 

PRINT

 key shifts the blinking digit. 

(5) Pressing the 

UNIT

 key counts up the blinking displayed value. 

(6) Repeats the steps (4) and (5) to set the weight value. 

(7) When aborting the setting, press the 

POWER/BRK

 key. Then 

” is displayed for several seconds, the balance stops   

the setting of weight value and returns to weight display. 

(8) After the setting of weight value is completed, press the 

TARE

 key.

(9) “

” is displayed for several seconds and the balance returns 

to weight display. 

(10) When setting the weight value exceeding the specified range, 

” is displayed and then the balance returns to weight   

display.  

 

 

Содержание CBL Series

Страница 1: ...2 ...

Страница 2: ...TIVITYCALIBRATION 17 1 Settingthevalueofsensitivitycalibrationweight 17 2 Sensitivity Calibration 19 REGISTRATION CANCEL ANDCHANGEOFUNIT20 SETTING 21 PCS No of pieces SETTING 22 PERFORMANCE CHECKS 23 MAINTENANCE 24 TROUBLESHOOTING 25 SPECIFICATIONS 26 PARTS LIST 27 PERIPHERALDEVICES 28 1 TheEP 60AElectronicPrinter 28 2 TheIFB 102ARS 232CInterface 29 3 Input outputFormat 31 4 CommandCode 32 ...

Страница 3: ...4 PRECAUTIONS ...

Страница 4: ...5 ...

Страница 5: ...n Pan supporter Instruction manual Inspectioncertificate Breeze guard Only type CBL120H CBL 220HandCBL320H Balance body Pan Pan supporter Pan shaft Level gauge Power jack Level screws 3 places Protection cover AC adapter option Data I O connector Breeze guard Only CBL120H CBL220H and CBL320H Breeze guard cover Only CBL120H CBL220H and CBL320H ...

Страница 6: ...N POWER BAK Selects Operation Warm up Cancels calibration or menu CAL MENU Performs calibration Selects a menu TARE Clears the displayto zero Sets a menu UNIT Selects a unit Sets or No of pieces Sets a value of sensitivitycalibration weight PRINT Outputs the displayed value to a printeror ortherextemal equipment Sets a value of sensitivitycalibration weight ...

Страница 7: ...ricshockwhenconnectingthepower usea3Pplugwithagrounding lineora2Pplugafterconnectingthegroundingwiretothescrewatthebackofthecase DonotplaceanythingthatmakesitdifficulttopulltheACadapterofftheoutlet Site Avoidinstallingthebalanceinaplacewhere Itisexposedtocorrosivegasorflammablegas Itisexposeddust wind vibration electromagneticwaves oramagneticfield Itisexposedtodirectsunlightorasuddenchangeintempe...

Страница 8: ...ightfrontandleftfront Finally inordertomakethebalancestable adjusttherightback screwtotouchthefloor Thisprocedureallowsyoutolevelthebalancequickly 3 Placethepansupporteronthepanshaftandthenthepanonthe pansupporter 4 Plug the AC adapter into the outlet The balance shows afterself checking 5 Press the POWER BRK key All displays light for one minute Then the display automatically shows zero and the b...

Страница 9: ...e POWER BRKkeywithoutdisconnectingtheACadapter Ifthebalanceisnotusedforonemonthormore disconnecttheACadapter CAUTION NOTES putwater metalpinoranythinginthebalance openthebalancecase leaveanythingexceedingtheweighingcapacityonthepan exposethebalancetoanythingmagnetized connectanythingotherthanthespecifiedequipmenttotheconnectorontherearsideofthe balance and giveashocktothepan ...

Страница 10: ...e Measurement 1 Whenusingatare loadthetareonthepanandpressthe TARE key after a stability mark is lit 2 Checkthatthedisplayshowszero 3 Load a sample When the stability mark is lit read the display If the total mass of the sample and tare exceeds the weighing capacity willbedisplayed Tare Asamplecontainerorother Stability mark Lightswhenthedisplayedvaluefallswithinthestabilityband Whenthe loadchange...

Страница 11: ...esnoothersetting The menu in the CBL series consists of three classifications Basically press the TARE key to go to lower hierarchy and press the POWER BRK key to return to upper hierarchy Continuously pressing the POWER BRK key returns the display to weight display from each hierarchy with single operation Step 1 PresstheCAL MENUkeyduringtheweightdisplay 2 willbedisplayed 3 EverytimetheCAL MENU k...

Страница 12: ...thevalueofsensitivitycalibrationweight RefertoSection7 2 Settingthevalueofsensitivitycalibrationweight Entersecondhierarchymenu Input outputformatsettingmode RefertoSection6 2 Menuselectionwhen isselected Weight display If the measurement is done at severe measurement environment and the stability of the display is notsogood setthebalanceto high stabilitymode When the high speed sample pouring mod...

Страница 13: ... key when the is displayed at 1st hierarchy menu enters 2ndhierarchymenu Inthismenu ON OFFofzerotracking settingthestabilitydetectionband registration cancelofunit ON OFFofautoprint andON OFFofanalogdisplayaremade Keyoperationandeachsettingaremadeasfollows ...

Страница 14: ...whenthedisplaystays within 1unit theresolutionvalueofthebalance forafixedperiodtime When the display shows the currentlysetconditionsaredisplayedon Zerotrackingeliminateszerodrift andshouldbeon for normal weighing Whenmeasuringweightchangesovertime orwhenslowlyaddingaliquidorpowdertothe balance turnoff thezerotrackingfeature ...

Страница 15: ...tcanbeset Baud rate 300 Baud rate 600 Baud rate 1200 Baud rate 2400 Baud rate 4800 Baud rate 9600 Delimiter C R Delimiter L F Delimiter C R L F Delimiter Comma Parity even Parity odd Parity none Stop bit 1 Stop bit 2 Press the TARE key Press the POWER BRK key Press the CAL MENU key When the display shows the currently set conditionsaredisplayedon ...

Страница 16: ...weighthavingknownvalue sensitivitycalibrationcanbemade Settheweightvalueusingatsensitivitycalibrationasfollows Step 1 Followingthemenuselection presstheCAL MENU key to make the display 2 PresstheTARE keytosetthebalancetoweightsetmode 3 Theweightvaluecurrentlysetisdisplayedandthedigittobeset blinks Whentheweightvalueisnotchanged presstheTAREkey 4 PressingthePRINTkeyshiftstheblinkingdigit 5 Pressing...

Страница 17: ... lessthan weighingcapacity CBL 620S More than 300g and less thanweighingcapacity CBL 1200H Morethan500gand lessthan weighingcapacity CBL 2200S More than 1000g and less thanweighingcapacity CBL 2200H More than 1000g and less thanweighingcapacity CBL 3200S More than 1500g and less thanweighingcapacity CBL 3200H CBL 3200HL More than 1500g and less thanweighingcapacity CBL 120H More than 50g and less ...

Страница 18: ...e calibration weight on the pan At this time the stability markwilloncedisappear 9 Whenthestabilitymarkislitagain presstheTAREkey 10 The display shows zero and blinks Make sure that the stability markislit 11 Unloadtheweight 12 Whenthestabilitymarkislitagain presstheTAREkey 13 is displayed for several seconds and the balance returns to weight display This completes the sensitivity calibration 14 I...

Страница 19: ... 4 PresstheTAREkeyontheunitdisplaytoberegistered Thatunitisregistered When three kinds of unit are already registered a new registration deletes the oldest registration among the three registered units However and pcs No of pieces cannot be registered simultaneously Thendeletesunnecessaryone 5 ContinuouslypressingthePOWER BRKkeyreturnsthedisplaytoweightdisplay Step for cancel 1 Carryoutthesameoper...

Страница 20: ...sasfollows dependingonreferencesampleweight REF Ifthe conversionisnotpossible isdisplayedfor severalsecondsandthebalancereturnstoweightdisplay H Type REF Minimumdisplayedvalue 100 conversionimpossible Minimumdisplayedvalue 100 REF Minimumdisplayedvalue 1000 100 Minimumdisplayedvalue 1000 REF Minimumdisplayedvalue 10000 100 0 Minimumdisplayedvalue 10000 REF Minimumdisplayedvalue 100000 100 00 Minim...

Страница 21: ...been alreadyregistered itisnotnecessarytoregisteragain 2 PlacethetareonthepanandpresstheTAREkey 3 Loadthestandardsamplewithrequiredpieces 4 Checkthatthestabilitymarkislit 5 WhenpressingtheUNITkeycontinuously thedisplaywillchange asfollows 6 SelectthedesiredPCSdisplayandpresstheTAREkey 7 isdisplayedforseveralsecondsandthebalanceenters thePCSunitdisplay Whenthereferencesampleweightislessthan readabi...

Страница 22: ...condition as follows Repeatability 1 Loadandunload10successivetimes an weight which is near the capacity of the balance Then record the following items Xi Displayedvaluewhentheweightisloadedafterstabilitymarkislit Yi Displayedvaluewhentheweightisunloadedafterstabilitymarkislit 2 Calculatethestandarddeviationof xand yusingtheformulasshownright Load X1 X2 Xi X10 Unload Y1 Y2 Yi Y10 Where Average val...

Страница 23: ...asfollows Pan Position of weight MAINTENANCE When fouled Whenthebalancebecomesdirty wipeitoffusingasoftclothwithasmallamountofmild detergent Organicsolventsorchemicaldustersshouldnotbeusedastheymaydamagepaintedsurfaces andthedisplaypanel Whenthebalanceisplacedinadustyordirtyenvironment usetheprotectioncoverofstandard accessory Thepancanbewashedwithwater Drythepansufficientlyandmounttothebalance 32...

Страница 24: ...onsite Change the measurement mode toHigh stabilitymode Influencefromelectricnoiseor electromagneticwave Maintainaproperdistancefromthe noisesource Zerotrackingworks Referto 6 MenuSelection Theaveragingprocessingisin High stabilitymode Changethemeasurementmodeifnecessary Therehasbeenaninstantaneouspowerfailure PressthePOWER BRK key the balanceentersweightdisplaymode Settingofcommunicationparameter...

Страница 25: ...0 250VAC 47 63Hz Balance 12VDC 0 1A Model CBL 120H CBL 220H CBL 320H CBL 1200H CBL 2200S CBL 3200H L Weighingcapacity 120g 220g 320g 1200g 2200g 3200g Readability 0 001g 0 01g Standarddeviation 0 001g 0 01g Linearity 0 002g 0 003g 0 02g 0 03g Calibrationweight 100g 200g 300g 1000g 2000g 3000g Pandiameter mm 100 100 Withtheguard 160 124 Mainbodysize mm Approx 170W 240D 114H Approx 170W 240D 75H Mai...

Страница 26: ...IMLF1class in the box 2kg 321 53449 ForCBL 2200S ForCBL 2200H 3200H L ForCBL 3200S Maintenance parts Partsname PartsNo Remarks Pan small For220S 320S 120H 220H 320H 321 54847 Pan Large For620S 2200S 3200S 1200H 2200H 3200H L 321 54846 Pansupporter small For220S 320S 120H 220H 320H 321 53908 11 Pansupporter Large For620S 2200S 1200H 2200H 321 53908 01 Pansupporter Large For3200S 3200H L 321 53908 0...

Страница 27: ...everthePRINTkeyispressed Autoprint Inthegdisplayofthebalance whenthedisplayiswithinzero 3count thedisplayisstabledwhen thesampleover20countsofgdisplayisloaded thebalanceautomaticallyprintsout Unloadthis sampleandwaitforthedisplayfallsintowithinzero 3counts thenloadanextsample Statistic calculation Pressing theSTATkey statistically calculates and prints the datauntilnextpressingoftheSTAT key SeeIns...

Страница 28: ...Ato the balance be sure to pull up theAC adapter for the balance TheninserttheplugofIFB 102AintotheDATAI Oconnectorofthebalance Signal PinNo Signal I O Function 1 FG Ground 2 TXD Output Dataoutput 3 RXD Input DataInput 4 RTS 5 CTS 6 DSR Input Transmittingispossiblewith polarity 7 SG Ground 20 DTR Output Receivingispossiblewith polarity ...

Страница 29: ...play valueofbalanceis beingdisplayedonthepersonalcomputescreenwheneverthe SPACE keyofpersonalcomputeris pressed Baud rate 1200 bps Parity none Delimiter CR means space IBM PC AT 10 OPEN COM1 1200 N 8 1 AS 1 20 Z INKEY 30 IF Z THEN 20 40 PRINT 1 D05 50 INPUT 1 A 60 PRINT A 70 GOTO 20 NEC PC 9801 VM2 10 OPEN COM N81NN AS 1 20 line and under the same IBM PC AT Setting baud rate 1200BPS by personal co...

Страница 30: ... Unit Unit At3 byte Unit Atplus Space Polarity Atminus Minus Atstable S Stabilityinformation onlyatoutputwithstability information Atunstable U For display U DL Atplus Space Polarity Atminus Minus Atstable S Stabilityinformation onlyatoutputwithstability information Atunstable U Data format ASCII JIS code Baud rate parity and delimiter change depending on menu selection ...

Страница 31: ... 60AElectronicPrinter D01 Continuousoutput Continuous output of data in the balance at every approx 100ms Forlessthan1200bps itbecomesapprox 150ms D09 Outputstop Autoprintorcontinuousoutputiscanceled D07 Singleoutputwithstability Information Printingismadeoncewithstabilityinformation D03 Continuousoutputwith stabilityinformation Continuousprintingismadewithstability information Q ON OFFselection T...

Страница 32: ...33 MEMO ...

Страница 33: ...34 MEMO ...

Страница 34: ...35 ...

Страница 35: ...36 ...

Отзывы: