background image

MODEL 7400-25 

INST-PM-7400-25 

Содержание 7400-25

Страница 1: ...MODEL 7400 25 INST PM 7400 25 ...

Страница 2: ...Revision Log Revisions Description Letter Date A 12 14 2016 Initial release B 04 04 2018 New cover ...

Страница 3: ...Kit for Model 7400 25 are shown below Note Photos may have a different appearance depending on chamber options 1 Turn off chamber and unplug power cord 2 Remove bottom shelf to expose humidity pan INSTRUCTIONS FOR REPLACING ANTIMICROBIAL COPPER ROD CO2 IN LINE FILTER AND AIR PUMP IN LINE FILTER IN MODEL 7400 25 REPLACING ANTIMICROBIAL COPPER ROD Air Pump In Line Filter CO2 In Line Filter Antimicro...

Страница 4: ...microbial copper rod as shown below 4 Discard old antimicrobial copper rod 5 Thread spring clip through cable loop slightly squeeze spring clip together and insert long white end of new antimicrobial copper rod into spring clip as shown below 6 Reinstall shelf Cable Loop Antimicrobial Copper Rod Spring Clip Spring Clip Antimicrobial Rod ...

Страница 5: ... 3 screws in top access panel along front of chamber using Phillips screwdriver as shown below Note Retain screws 3 Remove top access panel by sliding it towards front of chamber until 3 tabs on each side are released as shown below REPLACING CO2 IN LINE FILTER Disconnect electrical power before removing top access panel Screw Locations Access Panel Tabs ...

Страница 6: ...n line filter into tubing assembly 7 Tighten 2 clamps Note Photos may have a different appearance depending on chamber options 1 Loosen 1 clamp as shown below Note Clamps may be colored different than shown 2 Remove and discard old air pump in line filter by twisting and pulling filter out of tubing assembly 3 Install new air pump in line filter into tubing assembly 4 Tighten 1 clamp REPLACING AIR...

Страница 7: ...pplicable or plug in power cord and turn on chamber 6 Reset CO2 filter alert by pressing Settings button on right side of Home screen as shown below 7 Press Maint button as shown below 8 Press Reset button for Replace CO2 filter as shown below 9 Press Home button as shown below Settings Button Maint Button Reset Button Home Button ...

Страница 8: ...6 1 Reinstall top access panel by sliding towards rear of chamber until 3 tabs on each side are engaged 2 Secure top access panel along front of chamber with retained 3 screws using Phillips screwdriver REINSTALLING TOP ACCESS PANEL ...