APPLICATOR REPAIR
All repairs should be made on a clean, flat surface. If a
vise is used to hold parts during service or repair, DO NOT
clamp onto plastic parts and always pad the vise jaws!
The following parts should be thoroughly packed with
dielectric grease (LSCH0009-00) leaving NO air space or
voids when assembling:
• All O-Rings (PTFE o-rings do not need lubrication)
• Needle Shaft Assembly
• Packing Tube
• Cartridge Assembly, Non-Adjustable
• Air Valve Rod Assembly
Apply sealant (7969-10) to the external threads of the
following parts when assembling:
• Nut Air Valve Retaining (78635-00)
• Cap, Air Valve (79317-00)
Equipment Required
• Special Multi-Purpose Wrench (19749-00)
• 3/32" Allen Wrench for Set Screws
• Screwdriver (blade)
• Dielectric Grease (LSCH0009-00)
• Sealant, Medium Strength (7969-10)
• Plastic or Wood Dowel Rod, 5/16" Diameter
CAUTION
!
TO REMOVE THE APPLICATOR
FROM THE WORK SITE
ALWAYS
remove the applicator from the work
site for service or repair!
DO NOT
use any silicone lubricants in order to
avoid paint defects.
WA R N I N G
!
Solventborne Applicators
1. Flush the applicator. (Refer to "Flushing Procedures"
in the "Maintenance" section.)
2. With a wrench, loosen and unscrew the fluid hose
nut. Remove the fluid hose assembly.
3. With a wrench, lock the air fitting in place and loosen
the air hose nut using a second wrench. Completely
unscrew the nut and remove the air hose assembly.
4. Using a 16mm (5/8") open-end wrench, loosen the
low voltage cable from the cable plug. Pull the cable
straight out of the plug assembly.
5. Remove the applictor from the work site.
Ensure the control unit power is
OFF
prior to any
maintenance.
Figure 17: Removing Cascade Applicator
AH-06-01-R17 (08/2019)
39 / 79
www.carlisleft.com
MAINTENANCE
EN