![Carestream HEALTH CS 9000 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/carestream-health/cs-9000/cs-9000_user-manual_2557895040.webp)
Acquiring a 3D Single-Volume Image for Adult and Pediatric Patients
5–10
ACQUIRING IMAGES
Preparing and Positioning Adult and Pediatric Patients
To prepare and position the patient, follow these steps:
1. Ask the patient to remove all metal objects.
2. Ask the patient to wear a lead apron with a thyroid collar. Ensure that the apron lays
flat across the patient’s shoulders.
3. Ask the patient to enter the unit. On the
Control Panel
, press and hold
to adjust
the unit to the height of the patient.
4. Ask the patient to do the following:
•
Stand up straight or sit on a stool.
•
Grip the lower handle on each side.
•
Position their feet slightly forward.
•
Relax their shoulders for full motion of the unit rotative arm.
5. Position the patient:
•
Ask the patient to open their mouth.
•
Ask the patient to bite gently and naturally into the bite block without joining the
incisors.
•
Center the upper incisors with the bite block.