The detector should be tested every two months. Power the detector for at least two hours in a normal
operation environment to ensure the detector functions correctly. Return the detector to storage once
normal functionality has been verified.
8.2
Transportation
The packing instructions for the detector are as follows.
Pack the detector with a special antistatic bag and seal.
Place the EPE damping foam on the both ends of the detector.
Put the detector into the packing box and make sure the X-ray exposure surface is placed in an
upward position (
).
Transportation and storage conditions should be established during sale and purchase. Keep the
detector dry, upright, and in a secure position during transportation.
9
Specifications
9.1
Detector
Table 13 Detector specifications
Item
Description
Model
CareView 1500Cwe
Purpose
General radiography
Scintillator
CsI (Cesium Iodide)
TFT/PD
14” × 17”
Image matrix size
3072 × 2560 pixels
Pixel pitch
140 μm
Effective imaging area
430 mm × 356 mm
Grayscale
16 bit, 65536 grayscale
Fill factor
0.65
Spatial resolution
3.57 line pair/mm
Dynamic range
~82 dB
MTF
≥60%..........(@ 1lp/mm)
≥17%..........(@ 3lp/mm)
DQE
≥60%..........(@ 0lp/mm)
≥15%..........(@ 3lp/mm)
Rated power supply
Wireless
Wired
DC 24 V, Max. 1.5 A
Powered by the battery
Powered by the power box using interface cable
Power consumption
Max. 36 W