![CareFusion IVAC P6000 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/carefusion/ivac-p6000/ivac-p6000_technical-and-service-manual_2557183088.webp)
Fitting and Replacement Guidelines
P7000, P6000, TIVA, TCI & TIVA 88/106
1000SM00012 Issue 8
Torque Guide
Always use the correct torque level when making an assembly stage.
Take care with the torque applied when re-assembling parts.
The head patterns of the fasteners are of the following types:
- Pozi Number 1 (smaller X head)
- Pozi Number 2 (larger X head)
- M3 (Hex head with 5.5mm across flats (AF) drivers)
Always select the correct tool and bit pattern for the fastener.
The following list outlines the torque levels established during product manufacture.
Torque levels selected apply throughout product life for the IVAC® Syringe Pumps.
Use the information below as a guide to the 'do not exceed' torque levels when servicing the pump. When servicing, it is recommended
that torque is applied gradually until the component is secure. In any process do not exceed the stated levels.
If a torque driver is available for servicing this will help control the applied torque; otherwise, be aware that excess force may cause the
component to fail. If torque level is not stated then fixing should be hand-tight.
Transmission Assembly
Stage Description
Component Description
Qty
Established Process
Torque
Plunger holder to outer tube
Screw M3x8 Torx T6 Set Part Dog
2
45cNm
Carriage block to torsion rod
Screw M3x8 SKT Set Full
2
25cNm
Bonded Beam to transmission shaft
Hex Nut M3 Stainless Steel A4
1
25cNm
Earth leadscrew clip to transmission shaft
(Model TCI & TIVA)
Screw M2.5x8 CSK Head SKT A4 ST/ST
1
25cNm
Upper Case Assembly
Stage Description
Component Description
Qty
Established Process
Torque
Transmission assembly beam to Upper
Case
Screw M4x20 Csk Hd Posi SS
2
2.0Nm
Transmission assembly to Upper Case
Screw M4x8 Pan Hd Posi
3
1.0Nm
Screw M4x40 Pan Hd Posi 2 ZP+P
1
1.0Nm
Lower Case Assembly
Stage Description
Component Description
Qty
Established Process
Torque
RS232 Connector to Lower Case
Screw Male/Female M3.4 Brass & Nickel
2
25cNm
PSU to Lower Case
Screw M3x12 Pozi Hd Z+C
4
Hand tight
Adjustable Pole Clamp to Lower Case
Screw M4x16 Din 7985 ZP+P
2
20Nm
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...