Specifications
P7000, P6000, TIVA, TCI & TIVA 73/106
1000SM00012 Issue 8
The following information is for reference purposes only. For more detailed specifications refer to relevant
DFU
. Specifications refer to
all models covered by this manual, unless otherwise stated.
Infusion rate range
5ml syringes
0.1ml/h - 150ml/h
10ml syringes
0.1ml/h - 300ml/h
20ml syringes
0.1ml/h - 600ml/h
30ml syringes
0.1ml/h - 900ml/h
50ml + 100ml syringes
0.1ml/h - 1200ml/h
Note: Maximum infusion rate can be set as part of the configuration.
Volume infused
0.0ml - 9990ml
Bolus rates
5ml syringes
10 ml/h - 150 ml/h
10ml syringes
10 ml/h - 300 ml/h
20ml syringes
10 ml/h - 600 ml/h
30ml syringes
10 ml/h - 900 ml/h
50ml + 100ml syringes
10 ml/h - 1200 ml/h
Note: Maximum bolus rates can be set as part of the configuration. Bolus rates are
user adjustable.
Bolus Limit
25.0ml
Purge rate
100ml/h-500ml/h. Limited to maximum rate for syringe.
Purge limit
2.0ml
End Of Syringe, Keep Vein Open (KVO) rate
(P6000/P7000 only)
Stop, KVO (1.0ml/h), or set rate if lower than KVO.
Volume To Be Infused (VTBI) rate
(P6000/
P7000 only)
0.1ml - 100ml, 1 min - 24 hours
VTBI complete KVO rate
(P6000/P7000 only)
Stop, KVO (1.0ml/h, 2.0ml/h), set rate if lower than KVO, or continue at set rate.
Near End of Infusion (NEOI) Alarm
2.0% - 10.0% of syringe volume (TCI & TIVA/TIVA).
1 - 15 mins (P6000/P7000).
End Of Infusion (EOI) Alarm
0.5% - 5% of syringe volume.
Critical volume
0.5ml is the maximum-over infusion which can occur in the event of a single fault
condition.
Maximum Pumping Pressure Limit
280mmHg - nominal at L-3: factory preset (TCI & TIVA/P6000/P7000)
465mmHg - nominal at L-5: factory preset (TIVA)
650mmHg - nominal at L-7: highest alarm level
Occlusion levels L-0 to L-7.
System Accuracy
Drive Linearity
+/- 1%
Volumetric Mean
+/- 2% (nominal)
Important: System accuracy is +/-2% typical by volume as measured using the
trumpet curve test method defined in EN60601-2-24 at rates of 1.0ml/h and above
when the instrument is used with the recommended syringes. Differences in
factors such as size and plunger force in compatible syringes can cause variations
in accuracy and trumpet curves.) Also see trumpet curves section.
Infusion
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...