Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 69/106
1000SM00012 Issue 8
Procedure
1. Remove the two screws and washers securing the pressure transducer.
2. Remove the pressure tranducer and cable connector.
3. Reassemble in reverse order.
Pressure Transducer (Model P7000)
Spare Parts
Item
Description
Part Number
A
ASSY CONN PRESSURE SENSOR P7000
7000EL00014
B
SCREW M4X8 PAN HD POSI
0000ME00246
C
WASHER M4 WAVEY SST
0000ME00045
D
TRANSDUCER PRESSURE
147368-100
(
A
) Pressure sensor connector
(
B/C
)
Screws (x2), Washers (x2)
(
D
) Pressure transducer