Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 66/106
1000SM00012 Issue 8
Recommended when serviced
Bond syringe clamp if the pump was manufactured prior to July 1999 and serial number is within the
ranges
6001-00100 to 6001-00972
6002-00100 to 6002-00259
7001-00106 to 7001-04432
Procedure:
1) Remove the two circlips and spirol pin that secure the syringe clamp assembly in the upper case
then slide the syringe clamp assembly out through the bush in the upper case. The potentiometer
assembly and actuator can be left in place.
2) Carefully knock out the syringe shaft spirol pin, clean out the syringe shaft hole and apply Loctite
603, or equivalent adhesive around the edge of the hole.
3) Reassemble the syringe clamp assembly, re-inserting the pin through the hole in the clamp and
shaft.
4) Wipe off excessive adhesive and allow to cure for a minimum of 3 hours before refitting the
syringe clamp assembly into the pump. Note: If cure time of 3 hours cannot be tolerated,
activators may be used to reduce the cure time.
Syringe Size Pot, Syringe Clamp (continued)
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...