Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 65/106
1000SM00012 Issue 8
Syringe Size Pot, Syringe Clamp
Spare Parts
Item
Description
Part Number
A
SPRING 7.8 OD 44 LONG
0000ME00266
B
PLATE POTENTIOMETER 50MM PUNCHED
7000ME00063
C
SCREW M2X3 CSK HD SLOTTED
0000ME00164
D
PIN TENSION DIA 3.0X16MM
0000ME00116
E
CIRCLIP E TYPE SHAFT DIA 4.8
0000ME00002
F
ACTUATOR POT MOULDED SYRINGE SIZING
7000ME00033
G
ASSY POT P7000
7000SP00003
H
SCREW M3X6 PAN HD POSI ZP+P
0000ME00221
I
WASHER M3 WAVEY SST
0000ME00015
J
SHAFT SYRINGE CLAMP 100ML SYRINGE SIZING
7000ME00060
K
CLAMP SYRINGE MACH. V4
1000ME01006
L
PIN TENSION 3.0x10.0
0000ME00257
(
K
) Syringe clamp
(
A
) Spring
(
B
) Potentiometer plate
(
C
) Screw
(
D
) Pin
(
E
) Circlip
(
F
) Actuator
(
G
) Assembly pot
(
H
) Screw
(
I
) Washer
Procedure
1. Remove the potentiometer mounting plate fixing screw, remove pin which locks the syringe sizing pot and syringe
clamp in place.
2. Slide actuator off the syringe clamp shaft and remove the syringe sizing pot.
3. Reassemble in reverse order. See important note on next page.
When refitting, position the assembly pot so that actuator pot moves correctly when the syringe
clamp is moved up and down.
(
J
) Syringe clamp shaft
(
L
) Pin/Adhesive loctite 603
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...