Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 51/106
1000SM00012 Issue 8
TCI Diprifusor PCB, PE Stud Connector
Procedure
1. Remove the five screws which secure the cover plate to the underside of the lower case. Note: Use star driver to
remove fifth screw. Hinge the cover plate open and remove it.
2. Disconnect ribbon cable from the TCI Diprifusor PCB.
3. Remove the two screws which secure the TCI Diprifusor PCB and remove the PCB from the lower case, carefully pulling
through the aerial lead.
4. Disconnect the earth cable from the PE stud connector, as shown in Figure 2. below.
5. Remove the PE stud washers, plates and spacer slide out connector assembly.
6. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
ASSY PCB TIVA & TCI
6000EL00004
B
SCREW M3X6 PAN HD POSI ZP+P
0000ME00221
C
FLAT WASHER NYLON M3 TO ISO 7089
0000ME00044
D
STUD PE CONNECTOR M6 THREAD x 15
0000ME00141
E
LABEL SWITCH PANEL END TCI
6000LB00010 (see also 'Labels' section)
F
PLATE END CHASSIS TCI
6000ME00020
Disconnect
earth cable from
the PE stud
Figure 2. Earth cable
(
A
) TCI Diprifusor PCB
Aerial lead
Ribbon cable
(
E
) Label
(
D
) PE stud connector
(
F
) Plate end chassis
(
B/C
) Screws (x2) / Washers (x2)
Figure 1. TCI Diprifusor PCB Procedure
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...