Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 50/106
1000SM00012 Issue 8
RS232/Nursecall PCB (Mark I)
Procedure
1. Remove the four screws which secure the cover plate to the underside of the lower case. Hinge the cover plate open
and remove it.
2. Disconnect the RS232 ribbon cable from the Control PCB.
3. Disconnect cables from the RS232/Nursecall PCB. Remove the RS232 ribbon cable by pulling it through the slot in case
wall.
4. Remove the two nuts which secure the RS232 connector. Slide out the connector assembly and gasket.
5. Remove the three screws and washers which secure the RS232/Nursecall PCB and remove the board from the lower
case.
6. Reassemble in reverse order.
Refitting notes:
1) Apply loctite (243) to the threads of the RS232 screws (B) before assembly.
Spare Parts
Item
Description
Part Number
A
SPARES KIT RS232 OPTION
1000SP00210
B
SCREW MALE/FEMALE M3.4 BRASS & NICKEL
0000ME00279
(
A
) RS232/Nursecall PCB
(
A
) RS232 Ribbon cable
(
A
) RS232 Connector
(
A
) Gasket
(
A
) Screws (x3), Washers (x3)
(
B
) Screws (x2) /
Loctite 243
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...