Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 45/106
1000SM00012 Issue 8
Lower Case Assembly
Replacing the lower case
1. To replace a lower case, it will be necessary to fully strip down the old case and insert all the components into the new
lower case. This task requires a good knowledge of the pump, so be certain that you are fully conversant with all the
procedures in this chapter before undertaking this replacement.
2. To strip down each lower case subassembly, follow the instructions in the relevant section of this chapter. These
sections are:
Accessing the Pump
Battery
PSU, Alarm, Mains Inlet
Mains On/Off Switch (Mark I)
RS232 Connector
RS232/Nursecall PCB (Mark I)
TCI Diprifuser Module PCB, PE Stud Connector
Pole Clamp Assembly
3. To reassemble the components into the new lower case, simply reverse the order of disassembly.