Routine Maintenance
P7000, P6000, TIVA, TCI & TIVA 33/106
1000SM00012 Issue 8
Physical Inspection and Clean
To ensure the pump remains in good operating condition, it is important to keep it clean and carry out the routine
procedures described below. All servicing should only be performed by a qualified service engineer.
Thoroughly clean external surfaces of the pump, by wiping over with a cloth lightly dampened with warm water and a
standard disinfectant/detergent solution.
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
Recommended disinfectants are:
Brand Concentration
Hibiscrub
20% (v/v)
Virkon
1% (w/v)
Check that labels are flat, legible and fully adhered. Replace as necessary.
Inspect case components for damage and replace if necessary.
Inspect the pole clamp for damage and check that it functions correctly.
Inspect the AC power supply plug and cable for damage.
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow fluid to enter the
casing and avoid excess fluid build up on the pump.
Do not use aggressive cleaning agents as these may damage the exterior surface of the pump.
Do not steam autoclave, ethylene oxide sterilise or immerse this pump in any fluid.
Service Access Options (376)
Information logs and additional service options are available from the
SERVICE ACCESS
menu.
Use access code
376
to display the
SERVICE ACCESS
menu. See 'Access Codes' at the beginning of Chapter 2, 'Configuration
and Calibration' for further information on access codes.
Option
Description/Action
VIEW SERVICE LOG
Displays the last 10 fault codes. The time and date of each fault is shown.
SERVICE DATE
Set the date when ‘Service Due’ is required to be displayed on the pump.
SERVICE MESSAGE
Enter (or amend) a message to be displayed on the SERVICE DATE.
USE LOG
Displays the hours of use since cleared. Press CLEAR to reset hours to zero.
SERIAL NUMBER
Record the serial number of the pump.
EVENT LOG
Displays the event log (maximum 800 events). The time and date of each event is shown.
Note that pressing the
D
button on the pump at any time also allows access to the EVENT LOG option (if
enabled).
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...