16
Chapter 4 Unpacking and Setup
675–101–101 Rev. P
Figure 6: Flow pressure nomogram for 675-CFG-XXX when used with the Infant Flow LP or
AirLife variable-flow generator
Individual devices have a tolerance of up to ±15 percent from that illustrated in the
nomogram and, in particular, at pressures below 2cmH
2
O.
Flow Pressure Nomogram (for
reference only) to show typical flow
pressure relationships. This is not
meant to establish actual product
performance.
Содержание Infant Flow SiPAP
Страница 1: ...Infant Flow SiPAP Operator s Manual ...
Страница 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Страница 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Страница 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Страница 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Страница 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Страница 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Страница 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Страница 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Страница 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Страница 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Страница 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...
Страница 78: ...64 675 101 101 Rev P Infant Flow SiPAP ...