26
Servicing
If your unit requires service or repair please see page 37 for contact
details. A full service manual including circuit diagram and parts list
is available upon request.
Trouble Shooting Information
Should you encounter problems operating your BabyCO meter,
please consult the table below:
Problem
Possible Cause
Solution
BabyCO cannot be
switched on or “bat”
is displayed
Batteries are flat
Change the battery
Battery life is
shorter than
expected
Unit not being
switched off
Turn the unit off after
use
“CEL” is displayed
Fuel cell is
depleted
Return unit for cell
replacement
“Err” is displayed
Gas cylinder empty Check valve on cylinder
for contents and replace
cylinder if necessary
Check the calibration
value on the cylinder is
20ppm.
“bt2” is displayed
Internal battery
has expired
Return unit for battery
replacement
“gAS” is displayed
Fuel cell is
depleted
Return unit for cell
replacement
Fuel cell is
contaminated by
residual gas or
solvents
Allow unit to be
exposed to clean air.
Turn unit ON and await
message to disappear,
if not return unit for cell
replacement.
Содержание BabyCO Meter
Страница 2: ......
Страница 10: ...8 ...
Страница 41: ...39 ...
Страница 42: ...40 ...
Страница 47: ...45 6 5 4 1 2 3 CO FCOHb OFF ...
Страница 50: ...48 Das BabyCO ist zur Verwendung durch Ärzte und medizinisches Fachpersonal vorgesehen ...
Страница 79: ...77 ...
Страница 80: ...78 ...
Страница 114: ...112 ...
Страница 146: ...144 ...
Страница 181: ...179 ...
Страница 182: ...180 ...
Страница 216: ...214 ...
Страница 221: ...219 6 5 4 1 2 3 CO FCOHb OFF ...
Страница 254: ...252 ...
Страница 264: ...262 将滑动开关推至 CO PPM 位置 等待设备显示清零 如下所示 将控制阀紧紧拧到气缸上 并连接气源 串行 定标 开关 ...