1000DF00480 Issue 2 19/48
Multi-Step
The Multi-Step feature allows a sequential drug delivery program (up to 9 steps) to be set, delivering volumes of fluid at different
rates during each step. This allows the pump parameters to be set up once and to deliver a sequence eliminating the need to
change the rate and VTBI after each infusion step.
The infusion may be programmed in either rate and volume or volume and time. At completion of the last programmed step,
the channel switches to the preset KVO rate or remains at the current rate, whichever is less.
1. Select desired channel, as necessary. Channel must be on hold in primary mode.
2. Press
OPTIONS
key.
3. Press
Multi-Step
soft key.
4. Press
Enter New Program
soft key.
5. Select either
Rate and Volume
(pump calculates step infusion time) or
Volume and Time
(pump calculates rate).
6. To enter values for rate, volume or time as applicable, use numeric keypad. Press
ENTER
key.
7. To approve all displayed parameters and advance to next step press
ok
soft key.
8. Repeat steps 6 & 7 for each additional step required.
9. When all steps have been entered and accepted, press
done
soft key.
10. To approve and advance through review page(s), press
ok
soft key.
11. To clear VI, if required, press
CLEAR
or
0
(zero) key. Press
ENTER
key.
12. To approve
STEP TOTALS
page, press
ok
soft key.
13. To start Multi-Step infusion program, press
run
soft key or
RUN/HOLD
key.
Making Changes During Multi-Step
The channel must be on hold to view or edit the steps in the program.
1. To place channel on hold, press
RUN/HOLD
key.
2. To return to review page(s), press
setup
soft key.
•
A tick mark (
y
) next to a step on review page(s) indicates it has not started.
•
Only steps having a (
y
) can be edited.
•
A step number in progress is highlighted.
3. To advance through review page(s) of program, press
ok
soft key.
4. To select a step for editing, press a soft key.
5. To select value for editing, press a soft key.
6. To enter new value, use numeric keypad. Press
ENTER
key.
7. To return to review page(s) when programming is complete, press
ok
soft key.
8. To approve review page(s) and
STEP TOTALS
page, press
ok
soft key.
9. To resume infusion, press
RUN/HOLD
key or
run
soft key.
Viewing Totals Remaining
Press
4
soft key.
Time and VTBI remaining in Multi-Step program display for a short interval.
Resuming an Interrupted Multi-Step
The channel retains its place in the program if the pump is turned off. The program can be restarted from step 1 or resume where
it left off.
1. Select desired channel, as necessary.
2. Select New Patient and Profile Options, as necessary.
Return To Multi-Step?
page appears.
3. Press
yes
soft key.
4. Press
Review/Resume
soft key.
5. To resume program from point of interruption, press
Continue Program
soft key or to restart program at beginning of
step 1, press
Restart Program
soft key.
6. Verify all settings are correct. If a change is required, see “Making Changes During Multi-Step”.
7. To approve review page(s) and
STEP TOTALS
page, press
ok
soft key.
8. To continue or restart program, press
RUN/HOLD
key or
run
soft key.
Quitting Multi-Step
The channel must be on hold or the last dose complete.
1. Press
menu
soft key.
2. To return to primary set-up page, press
Quit Program
soft key.
Options (continued)