Getting Started (continued)
1. To turn channel on, press channel’s
POWER
key.
•
The pump will run a short self-test.
•
Check the display test pattern and ensure that no rows or pixels are missing.
2.
Confirm Profile?
- Press
change
to change Profile, Profile selection menu displays, select profile and
ok
. Go to step 3.
Accept
will retain profile and display
NEW PATIENT?
screen.
A
MENU screen will show either Resume Previous Drug or Return to ....? if New Patient is not selected.
Option shown will be dependent on previous infusion.
3.
New Patient?
- Answering
Yes
will clear all previous patient data.
No
will retain previous patient data.
MENU
Screen
displays.
4.
MENU
screen - Select one of the following:
•
Primary Infusion
- allows infusions to be given when a drug is not listed or when a data set is not loaded.
•
New Guardrails Drug
- allows a drug to be selected from the Drug Library in the data set.
•
Previous infusion resume option, if available.
A
Before each use, verify all infusion settings are appropriate for the patient.
Primary Infusion
1. Enter rate value, use numeric keypad then press
ENTER
key.
2. To enter desired VTBI, use numeric keypad then press
ENTER
key.
3. If there is a VI value that needs to be cleared, press
CLEAR
key or press
0
(zero) key then press
ENTER
key.
4. Verify that all parameters are correct, then press
RUN/HOLD
key to start infusion.
5. To briefly view current profile from
RUN/HOLD
page or during a running infusion, press
4
soft key.
New Guardrails Drug
1. Select a drug name by pressing the adjacent soft key. When a drug is not listed select
Drug? NO DOSE LIMIT
.
2. Select Concentration, Dose units and weight or height unit, if applicable.
3. To confirm selection and continue programming, press
ok
soft key.
4. To enter or adjust a value, use numeric keypad then press
ENTER
key and use the softkeys to highlight parameters to
change. Press
ok
soft key to confirm.
5. To enter desired VTBI, use numeric keypad then press
ENTER
key.
6. If there is a VI value that needs to be cleared, press
CLEAR
key or press
0
(zero) key then press
ENTER
key.
7. To accept infusion parameters, press
ok
soft key.
8. Verify that all parameters are correct, then press
run
soft key to start infusion.
9. To briefly view setup parameters (concentration, patient weight, current profile) from
RUN/HOLD
page or during a
running infusion, press
4
soft key.
Resuming Previous Infusion
If a channel has been powered off during an infusion, previous programming may be resumed if:
•
One channel of a dual channel pump remained on.
•
Current profile is accepted and
New Patient? – no
is selected during start up
1. Select desired channel as necessary.
2. Press soft key next to the resume option required. Previous set-up screen displays.
3. Verify that all parameters are correct or change as required, then press
run
soft key or
RUN/HOLD
key to start infusion.