5B-24
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.20 Removing Motor/Pulley Assembly
1. Remove four (
4
) screws attaching
Motor/Pulley Assembly to Bottom Plate.
2. Remove Motor Belt.
3. Remove Motor/Pulley Assembly.
During Reassembly:
1. To prevent pinion damage, install
Motor/Pulley Assembly through Bottom
Plate at a slight angle.
2. Install Motor Belt:
a. Install wide end of belt onto
leadscrew pulley.
b. Tilt Motor/Pulley Assembly and
install other end of belt onto motor
pulley. Do not stretch belt to fit over
pulley.
3. Screw Motor/Pulley Assembly into place.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Motor Belt
Motor/Pulley Assembly