5B-17
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.14 Removing Syringe Size Sensor
Assembly
1. Remove one (
1
) screw attaching Syringe
Size Sensor Assembly and Shaft
Support to Front Case.
2. Remove Syringe Size Sensor Assembly.
During Reassembly:
1. Insert tab on Syringe Size Sensor
Assembly into Potentiometer Extension
Guide.
2. Secure Syringe Size Sensor Assembly
to Front Case.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Syringe Size
Sensor Assembly
Potentiometer
Extension Guide
Front Case