5B-11
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Internal Frame Assembly
5.2.8
Removing Logic Board Assembly
1. If present, remove Retainer Clip from
Dual Connector Cable-Logic Board
connection.
2. Disconnect Dual Connector Cable from
Logic Board
J8
.
3. Disconnect Linear Sensor Harness from
Logic Board
J12
.
4. Remove one (
1
) screw attaching Logic
Board to
IUI
Support Bracket.
5. Remove one (
1
) screw attaching Logic
Board to Internal Frame Assembly.
6. Disconnect harnesses from following
locations on Logic Board:
J4
(display)
J7
(motor)
J11
(flange detect)
J13
(lock harness)
7. Remove Logic Board.
During Reassembly:
Route Dual Connector Cable and Linear
Sensor Harness under
IUI
Support Bracket.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Dual Connector Cable
Logic Board
J12
Linear Sensor Harness
IUI Support Bracket
J8
J11
J4
J7
J13
Retainer Clip
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......