5B-9
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5.2.6
Removing Flex Cable Retainer
1. Peel Flex Cable Retainer from Internal
Frame.
2. Use a cotton swab or soft cloth lightly
dampened with
70%
Isopropyl Alcohol to
remove adhesive debris from Internal
Frame.
During Reassembly:
1. Ensure Internal Frame surface (where
retainer is adhered) is clean and free of
adhesive debris.
2. Remove protective backing from
replacement retainer and adhere to
Internal Frame.
3. If necessary, bend retainer on each fold
line at a
90°
angle, toward front of
instrument.
4. Lay retainer over Flex Cable and tuck
inside Front Case.
Flex Cable Retainer
Retainer tucks into
Front Case here.
Retainer adheres to
Internal Frame here.