CORRECTIVE MAINTENANCE
Syringe Module
5A-10
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.8
Removing Internal Frame Assembly
1. Remove three (
3
) screws attaching
Internal Frame to Front Case Assembly.
2. Remove
IUI
Bracket Strap.
3. Remove one (
1
) screw attaching
IUI
Support Bracket to
IUI
Bracket.
4. Remove
IUI
Support Bracket and
IUI
Bracket.
5. Remove two (
2
) screws attaching
Internal Frame to Drivetrain Assembly.
6. Remove Internal Frame.
During Reassembly:
Route Linear Sensor Harness between
IUI
Bracket and Internal Frame.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Drivetrain Assembly
IUI Bracket
Internal Frame
IUI Bracket
Strap
IUI Support Bracket
Front Case Assembly
Linear Sensor Harness
To replace an Internal Frame whose Barrel
Clamp Assembly has a park position “pin”
(rather than a “collar”), replace with Front
Case Assembly Kit (Table
9
).
CAUTION
-- Barrel Clamp Assembly
(Partial View)
--
Park Position “Collar”
Park Position “Pin”
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......