CORRECTIVE MAINTENANCE
Syringe Module
5A-4
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5.2.2
Removing IUI Connector Assemblies
Remove two (
2
) screws attaching each
IUI
(left and right) to module.
During Reassembly:
• Do not mix gray (old) and black (new)
IUI
Connectors.
• Ensure ground clips are still installed on
both
IUI
connectors.
• To install Right
IUI
Connector Gasket,
remove protective backing and adhere to
IUI
Connector.
• To install Left
IUI
Connector Seal,
position seal on one end of connector
and stretch to other end to conform to
connector body. Gently press on seal to
seat completely. Use lint-free swab to
apply alcohol to top, sides, and bottom of
seal for lubrication while installing to Rear
Case. Do not apply alcohol to contacts
or connector.
IUI Connector, Left
IUI Connector, Right
IUI Connector Seal, Left
IUI Connector Gasket, Right
Ground clip at each mounting
hole on both IUI Connectors.
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......