ILLUSTRATED PARTS BREAKDOWN
Syringe Module
7A-8
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Figure 7A-2.
Housing Assembly
1
Install 6
½”
Gasket (item
281
) around parameter,
and into undercut
feature, of Upper
Housing (item
285
).
Gasket is not to overlap
into opening for
Actuator Knob
Assembly (item
3
).
2
281 1
250
285
200
205
478
474
203
2
461
3
4
Replacement of the
Force Sensor Assembly
in a Syringe Module
with version
7
or earlier
software must be
performed at the depot.
further information.
4 PL
4 PL
4 PL
To avoid damaging Lower
Housing (part of item
4
), use
Nylok screws (item
474
) to
install Force Sensor
Assembly (item
250
).
CAUTION
3
Nylok screws (item
474
)
are provided as part of
Lower Housing/
Carriage Assembly Kit
(item
4
) and Force
Sensor Assembly (item
250
).
3
3
3
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......