5
Note!
•
When using a lift to place the user in the
bed and the head end of the bed is raised,
make sure that the user is not placed too
high on the mattress. Otherwise, there is a
risk of shear.
•
When the alternating or pulsating
program is used and the head end of the
bed is raised, make sure that the user
and/or the mattress is not moving
downwards due to the movement in the
mattress. Also raise the foot end of the
b d
3
Common operations
Following operations apply to CuroCell® M4,
CuroCell® A4 and CuroCell® IQ regardless of
which mattress that is used.
3.1
Sitting positioning in bed
When raising the head end of the bed into a
sitting position, always secure the user´s
position. To ensure the product functionality,
we always recommend to perform a hand
check (see section 4, 5 or 6 depending on the
pump and mattress of interest). This function
is recommended to use for short periods only.
For additional support, positioning pillows can
be used.
3.2
CPR (Cardipulmonary resuscitation)
In case of an emergency where CPR
(Cardiopulmonary resuscitation) is necessary,
remove the connection (marked ‘CPR’) from
the control unit
and leave the
lid open in
order to empty
the mattress of
air quickly.
3.3
Transport function
If the user needs to be moved in bed, either:
•
Unplug the CPR connection, close the lid,
place the CPR connection at the end of the
bed and remove the control unit from the
bed. The mattress will retain air for at least
12 hours.
Or
•
Remove the power supply from the wall
socket and leave the control unit hanging on
the bed during transport. The mattress will
retain air for at least 12
hours.
We recommend using this function for short
periods only.
3.4
Restart
If a restart is required, set the On/Off switch
on the side of the control unit to 0 (Off). Wait
for approx. 10 seconds and restart the control
unit.
3.5
Maximum pressure notification
When the function Maximum
pressure has been used for a long
time, the Maximum pressure diode
will blink. If the use is intentional, ignore the
notification.
3.6
Power failure
In the event of a power failure, unplug the CPR
connection, close the lid, place the CPR
connection at the end of the bed and remove
the control unit from the bed. The mattress
will retain air for at least 12 hours. Perform a
hand check to make sure the pressure of the
mattress is not too hard or too soft.