![Cardioline XR50+ Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/cardioline/xr50/xr50_operators-manual_2556434030.webp)
28
User Manual: xr50+_eng
Cardioline S.p.A.
Via Linz,151
38121 Trento (TN)
www.cardioline.com
Sales office:
Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470
fax +39 02 94750471
xr50+
9 Electromagnetic Compatibility EN 60601‑1‑2
Changes or modifications to this system not expressly
approved by ergoline GmbH could cause EMC issues with
this or other equipment.
This system is designed to comply with applicable regula‑
tions regarding EMC.
Its compliance with these requirements has been verified.
It needs to be installed and put into service according to
the EMC information stated as follows.
Warning
• RF Interference •
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
Caution
• Equipment Malfunction •
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The xr50+ ergometer is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that the xr50+ ergometer is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions to EN 55011
Group 1
The xr50+ ergometer uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby elec‑
tronic equipment.
RF emissions to EN 55011
Class B
The xr50+ ergometer is suitable for use in all establish‑
ments, including domestic and those directly connected
to the public low‑voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions to EN 61000‑3‑2
Class A
Voltage fluctuations / flicker emissions
to EN 61000‑3‑3
Complies