8400-M022-O1 Rev K
Model 190 Indicator
12
INDICATOR SETUP
Calibration Inhibit Jumper
Your Model 190 indicator has been thoroughly tested and calibrated before being shipped to
you. If you received the indicator attached to a scale, calibration is not necessary. If the
indicator is being connected to a scale for the first time or recalibration is necessary for other
reasons, proceed as indicated.
Calibration and Setup of the indicator are accomplished entirely by the keypad. However, it
may require changing the position of the calibration inhibit jumper depending on the method of
sealing required by your local metrology laws.
The calibration inhibit jumper (
J4
) is located on the main printed circuit board (see Figure
No. 8) and can
only
be accessed by removing the front panel of the indicator.
Figure No. 8
IMPORTANT!
The following setup parameters
CAN NOT
be changed with the
calibration inhibit jumper (
J4
) installed:
USA
= Domestic or International
LFt
= Legal-For-Trade
Unit1
= Weighing Units 1 (Primary Units)
Int
= Interval Setting
dPP
= Decimal Point Precision
CAP
= Capacity
Unit2
= Weighing Units 2 (Secondary Units)
trA
= Zero Tracking Range
trL
= 4% Zero Limit
PU0
= Power Up Zero
dFLt
= Digital Filter Number
F
= Filter Level Amount
B
= Filter Break Range
Sr
= Sample Rate
UnS
= Motion Range
SC
= Stable Count
J4
CAL Inhibit
Jumper J4