Carbolite HTR 11/150 Скачать руководство пользователя страница 7

 

HTR 

 

 
MF30 – 3.07 

4.0

 

M

AINTENANCE

 

4.1

 

General Maintenance 

 

No routine maintenance is required. The outer surfaces may be cleaned with a damp cloth. Do not 
allow water to enter the interior of the case, tube or control box. Do not clean with organic 
solvents. 

 

Silica vessels should be kept clean - see section 3.5. 

4.2

 

Calibration 

 

After prolonged use the controller and/or thermocouple could require recalibration. This would be 
important for processes which require accurate temperature readings or which use the furnace 
close to its maximum temperature. A quick check using an independent thermocouple and 
temperature indicator should be made from time to time to determine whether full calibration is 
required. Carbolite can supply these items. 

 

Depending on the controller, the controller manual may contain calibration instructions. 

4.3

 

After Sales Service 

Carbolite’s service division (Carbolite Engineering Services) has a team of Service Engineers 
capable of repair, calibration and preventive maintenance of furnace and oven products at our 

customers’ premises throughout the world. We also sell spares by mail order. A telephone call or 
fax often enables a fault to be diagnosed and the necessary spare part despatched.  

 

Each furnace has its own record card at Carbolite. In all correspondence please quote the serial 
number, model type and voltage given on the rating label of the furnace. The serial number and 
model type are also given on the front of this booklet when supplied with a furnace. 

 

To contact Carbolite Engineering Services or Carbolite see the back page of this manual. 

4.4

 

Recommended Spares Kits 

 

Carbolite can supply individual spares, or a kit of the items most likely to be required. Ordering a 
kit in advance can save time in the event of a breakdown. Each kit comprises one thermocouple, 
one sheath, one solid state relay, and one heating element. 

 

When ordering spares please quote the model details as requested above. 

4.5

 

Power Adjustment 

 

The furnace control system incorporates electronic power limiting. Depending on the model this 
has a maximum setting related to the supply voltage. The power limit parameter 

OP.Hi

 may be 

inaccessible to the operator.  

 

Occasionally the power limit is set to zero to permit demonstration of the controls without the 
heating elements taking power. In this case the power limit is accessible to the operator and may 
safely be reset to its standard value. See section 9.0 for details of power limit settings. 

Содержание HTR 11/150

Страница 1: ...onnected to the electricity supply Manuals are supplied separately for the furnace controller and overtemperature controller when fitted Please read the controller manuals before operating the furnace...

Страница 2: ...e control circuit is energised Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements DANGER of electrical shock read any warning printed by this symbo...

Страница 3: ...n the amperage on the label A table of the most common fuse ratings is also given in section 8 1 of this manual Where a supply cable is present there are internal supply fuses customer fusing is prefe...

Страница 4: ...and tighten part 1 with the spanner Do not overtighten The end of the vessel with this assembly fitted becomes the left hand end Lightly lubricate the drive coupling assembly and the furnace drive wh...

Страница 5: ...temperature controller according to the instructions supplied To turn the furnace off set the Instrument switch to it s off position the controller display will go blank If the furnace is to be left o...

Страница 6: ...ature This is a recrystallisation process which is greatly affected by surface conditions and therefore by contaminants in the furnace atmosphere At 1000 C only a highly contaminated atmosphere has a...

Страница 7: ...all or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carbolite In all correspondence please quote the serial number model ty...

Страница 8: ...urther information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Replacement 301 This controller is fitt...

Страница 9: ...no vessel in place Make a careful note of the colours and positions of all the electrical connections to the inner cylindrical chamber and thermocouple Disconnect the thermocouple and element connecti...

Страница 10: ...or faulty controller There are no lights glowing on the controller Check the supply fuses and any fuses in the furnace control compartment The controller may be faulty or not receiving a supply due to...

Страница 11: ...Reactor 208V 240V Instrument Switch coil temperature controller thermocouple overtemp controller thermocouple SSR e l e m e n t s heat on F1 F2 N L PE Filter if fitted contactor or relay overtemperat...

Страница 12: ...trument Switch coil temperature controller thermocouple overtemp controller thermocouple SSR e l e m e n t s heat on F1 F2 N L PE Filter if fitted contactor or relay overtemperature if fitted motor co...

Страница 13: ...HTR MF30 3 07 13...

Страница 14: ...n board 20mm x 5mm Other 32mm x 6mm F3 Heat Light fuses May be omitted up to 25 Amp phase supply rating 2 Amps glass type F 32mm x 6mm Customer fuses Required if no supply cable fitted Recommended if...

Страница 15: ...e the setting beyond the value in the table the wiring and the elements are designed on the basis of the declared settings Model Volts 200V 208V 220V 230V 240V 110V 115V 120V HTR 11 75 100 100 89 81 7...

Страница 16: ...n of all Furnace and Oven products please contact Carbolite Engineering Services Telephone 44 0 1433 624242 Fax 44 0 1433 624243 Email service carbolite com MF30 3 07 14 08 14 9 0 SPECIFICATIONS Carbo...

Отзывы: