13.0
Decommissioning, Storage and Disposal
13.1
Decommissioning
1. Reduce the setpoint on all temperature controllers to 0°C so that heating cannot
begin accidentally when the product is recommissioned
2. Isolate the product from the power supply
3. Allow the product to cool to room temperature
4. Disconnect the product from the power supply
5. Disconnect all additional equipment and external connections e.g. gas supplies
13.2
Storage (Long Term)
Store in a cool, dry place.
Note:
If the furnace is exposed to a humid environment during storage,
it must be fully dried out before recommissioning. All internal electrical
circuits should be checked for signs of moisture. If there are visible
signs of moisture, the product should be isolated from the power supply
and allowed to dry out at ambient temperature for at least 24 hours.
Please contact Carbolite Gero Service for further advice.
13.3
Disposal
Note:
This product should only be disposed of in accordance with
local regulations and requirements regarding electrical equipment.
Within the European Community the disposal of electrically operated devices is
regulated according to guidance based on the EU Directive 2012/19/EU on Waste
Electrical and Electronic Equipment (WEEE). Disposal regulations may differ worldwide.
If uncertain, please contact Carbolite Gero for advice on disposal.