CAPINTEC, INC.
CRC
®
-25R
Indicating Daily Usage
It must now be decided whether the source will be used for the Daily Test.
Co57
Use Daily ?
Yes or No
Figure 6-13 Daily Use
If the Test Source will be used in the Daily Test, press
YES
. If the Test Source will not be
used in the Daily Test, press
NO
.
Confirming Calibration Data
Figure 6-7 Test Source Verification Screen will now be displayed with the data that has just
input.
If the displayed information is not correct, press
NO
. Figure 6-8 Source Existence will
appear.
If the displayed information is correct, press
YES
. Figure 6-5 Test Source Menu will re-
appear.
CONSTANCY TEST SOURCE
The source for the Constancy Test is chosen from the Test Sources designated as Daily
Sources.
From the Figure 6-5 Test Source Menu, select
CONSTANCY
.
No Daily Source Chosen
If no Test Sources are designated as a Daily Source, Figure 6-14 No Daily Source Was
Chosen will appear.
6 - 8
CHAMBER INITIALIZATION
March 15
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......