CAPINTEC, INC.
CAPTUS
®
3000
as default settings each time a patient is entered into the RBC Survival Test patient
directory. These settings may be changed as needed for any individual patient or
patients.
Figure 10-65
Default Values
frame –
Counting Time
– This is duration for counting in seconds each
time a measurement is done. The default is set to 60 seconds.
Repetitions
frame – This is number of repetitions each time a measurement is made.
For two repetitions, the average value is used in calculations. The second repetition may
be skipped during the measurement procedure if so desired.
Normal Range for 50% Survival
frame –
Minimum
and
Maximum Days
– These
numbers will define a range for number of days over which 50% RBCs surviving can be
considered as normal. They are entered as seen fit by the doctors. Out of range results
indicate a medical condition and will be flagged in the report. Entering these values is
optional.
2. To enter new values in any frame, click in the appropriate text box or radio button or use
the
Tab
key to move from one selection to another.
3. Click the
button to save changes or click the
button to discard the
changes and return to Figure 10-64 RBC Survival Test Main Screen.
QUICK TIPS:
•
Counting Time
field appears again, while adding a patient for the RBC
Survival Test. Any changes made on the Add Patient screen will override the
default values entered in the Setup.
•
For greater flexibility, set
Repetitions
to 2, as the second repetition may be
skipped if desired, while performing the measurements.
•
To return to previous screen without saving changes press the
Esc
key.
Adding Patients
Note: All fields in bold are required before you can begin the RBC Survival Test. However,
you may enter the Patient Name and ID alone, click the
button and enter
10-44
LAB TESTS
November 13
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...