P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement

                                  

!

MicroBio MB2-RSH

!

!

Installation and Operation Supplement

!

!

!

!

!

!

!

!

Document No.

!

P0002W006

!

!

Issue:

!!

!

2

!

!

Date:

! !

!

23 July 2010

!

!

Issue 2

!

/

                                                                                                                                                          1 12

Содержание MicroBio MB2-RSH

Страница 1: ...ium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement MicroBio MB2 RSH Installation and Operation Supplement Document No P0002W006 Issue 2 Date 23 July 2010 Issue 2 1...

Страница 2: ...Cantium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement Issue 2 2 12...

Страница 3: ...n and Operation Supplement Document Issue 5 Technical Specification 6 Installation 6 Fitting the Connector for Unterminated Cables 6 Connecting the Control Unit and remote Sampling Head 10 Calibration...

Страница 4: ...Cantium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement Issue 2 4 12...

Страница 5: ...2W006 MicroBio MB2 RSH Installation and Operation Supplement Document Issue Issue Description Date Revision by Approved 1 First release 20 June 2010 DCP SNJP 2 Added section on bio decontamination 23...

Страница 6: ...e other with one of the connectors not fitted By not terminating one of the connectors enables the cable to be passed through glands or bulkheads as required by the end user Once the cable is routed t...

Страница 7: ...ic Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement 2 Fit connector housing and strain relief onto cable end 3 Remove cable clamp 4 FIt frame onto cable end and refit cable cla...

Страница 8: ...o connector pins The connector body has pin numbers embossed onto the surface adjacent to the pins i Pin 1 Red ii Pin 2 Blue iii Pin 3 Screen 6 Fit pin housing back into frame noting that there are ra...

Страница 9: ...icroBio MB2 RSH Installation and Operation Supplement 7 Slide the assembly strain relief and metal shell together 8 Align shell so that the brass thread of the frame body can be seen then replace the...

Страница 10: ...is to the extension cable end labelled RSH The connectors have a screw locking mechanism Screw the two halves together by twisting the outer ring of the plug from the the remote sampling head connecto...

Страница 11: ...nsion cable from the control unit and then plug the remote sampling head directly into the control unit as below If extension cables over 10m in length are used then it is recommended that the sampler...

Страница 12: ...1000 ppm Conditioning time 20 minutes Gassing time 30 minutes Aeration time 150 minutes As the MicroBio MB2 RSH has a fan within the remote head it is recommended that during the decontamination proce...