3.4 Display on Measurement Screen
– 8 –
3.4 Display on Measurement Screen
The screen below is an example when FULL AUTO mode is selected.
3.4.1 Before Measurement
F U L L AU TO ( 3 )
L I M I T E R O F F
N o . 0 0 0 0 0 1
mmHg
STD
[ , , ]
L
[ , , ]
3 0
R
Right or left side eye
The display of the side of the eye
where the measurement head is
positioned will be reversed.
Condition of limiter
When the limiter for preventing the
nozzle from coming into contact
with examinee’s eye is not working,
message “
LIMITER OFF
” will be dis-
played.
Limiter can be set to ON/OFF with
LIMITER switch. See section 5.1.
Serial number
Serial number is displayed when “
COUNT
”
on page 1/3 of SET mode is set to “
ON
”.
The initial value can also be set from the
same page of SET mode.
Measurement mode
Measurement mode (FULL
AUTO, AUTO or MANUAL) is
displayed. Measurement mode can
be selected with AUTO/MANU.
switch.
Number of times of measurement
The number of times of measurement (1 or 3) to be per-
formed in FULL AUTO mode is displayed.
It is not displayed in AUTO and MANUAL mode.
Number of times can be selected by pressing PUFF
TIMES switch.
Order of displaying the result
Order of displaying the result (“ ”
for reliability or “
” for sequence)
is shown. Order can be selected
from “
ORDER
” on page 1/3 of SET
mode.
Measurement range
Measurement range (“
30
” for 0 –
30 mmHg or “
60
” for 25 – 60
mmHg) is displayed. It can be
selected from “
RANGE
” on page
1/3 of SET mode.
Unit of intra-ocular pressure
Alignment ring
This is the ring used for
positioning the examinee’s
pupil during manual mea-
surement.