CANNON Instrument Company | Software Menu Options
31
Name
Description
Status bar
Displays or hides the status bar, which displays information about
the current application status or other contextual information
generated by the location of the cursor or mouse icon.
Instrument status
Displays or hides the instrument status window, which displays
current force, deflection and temperature information, as well as the
target BBR temperature
Graph*
Opens a graph view of the current BBR data file
Report*
Opens a columnar report view of the current BBR data file
*option available only when test file is open
Table 6: Calibrate Menu Options
Name
Description
Force, deflection and
compliance
Calibrates the Load Unit for maximum accuracy (see Calibration for
more information)
Thermistor calibration
Calibrates the thermistor probe to insure accurate temperature
readings (see Calibration)
Auto-Thickness
Calibrates the BBR for precision measurement of specimen thickness
using the thick steel beam as a reference standard.
Run a Confidence Test
Initiates the steel beam test sequence to verify normal operation of
the BBR instrument. Run a Confidence Test verifies the function of
the BBR instrument using a procedure which measures the known
stiffness modulus of a steel beam (see Calibration
Setup Menu Options
Instrument Setup: The
Instrument Setup
menu provides access to important software/system
configuration settings.
Set Bath Temperature: Bath temperature is also set from the
Instrument Setup
menu. To enter the
desired bath temperature:
1.
Click
Bath
from the tab options on the
Instrument
Setup
property sheet.
2.
Enter the desired temperature (b25 °C and -40 °C).
3.
Click
OK
to save the settings, or
Cancel
to discard the changes.
Содержание TE-BBR
Страница 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Страница 2: ......
Страница 10: ......
Страница 11: ......
Страница 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Страница 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Страница 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Страница 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...