79
CANNON
®
Mini-Rotary Viscometer
CMRV-5000
Instruction & Operation Manual
Version 1.0b—August, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
Cooling Profiles
The cooling profile may be pictured graphically as a curve (temperature
over time) calculated from up to 25 individual user-defined data points.
To design a profile, you will use the Profile Editor from the VISCPRO
®
Profile Designer window to input
time
,
temperature
and (if desired)
tolerance
values for each point:
Time:
The time during the sample-run at which a desired temperature is to be
achieved. Acceptable values for hours are 0-99. Acceptable values for
minutes are 0-59. Note that the first point in a profile should NOT have
a null value of 0 hours AND 0 minutes.
Temperature:
The required CMRV temperature at the particular time specified.
Acceptable parameters are +90°C to -40°C.
Tolerance:
The acceptable CMRV temperature variance (to the nearest 0.1°) from
the specified temperature value (in °C). This field may be left blank if no
tolerance is required.
The Profile Designer
Cooling profiles are displayed and maintained using the VISCPRO
®
Profile Designer window. The Profile Designer window provides a user-
friendly interface for selecting and/or creating profiles and, if necessary,
modifying them to create additional profiles.