1-12
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Optical Precautions
Identify the laser room clearly. Post appropriate warning signs in prominent locations at
all entrances to the laser room.
Cover all windows, portholes, etc. with opaque material to prevent laser light escaping
from the laser room.
Restrict entry to the laser room when the AlexTriVantage Laser is in operation. Allow
access to the laser room only to those personnel both essential to the procedure and
well trained in laser safety procedures.
Make sure that all laser room personnel are familiar with the laser system controls and
know how to shut down the laser system instantly.
Appoint one person to be responsible for the laser system controls during the
procedure.
Ensure that all personnel wear appropriate safety eyewear whenever the laser system
is on.
Avoid accidental exposure to the laser beam either directly, or reflected from a surface,
by ensuring that all personnel wear appropriate safety eyewear whenever the laser
system is on. Verify that the protective eyewear used is known to protect against the
wavelengths emitted by the AlexTriVantage Laser.
Never look directly into the laser beam coming from the laser system, or reflected from
a surface, even when wearing protective eyewear.
Never allow the laser beam to be directed at anything other than the targeted area, the
calibration port or a safe beam stop.
Never permit reflective objects such as jewelry, watches, instruments or mirrors to
intercept the laser beam.
When the laser system is not actually being used, place the AlexTriVantage Laser in
the Standby state. This will prevent accidental pulsing of the laser should anyone
inadvertently depress the trigger switch.
Never leave the key in an unattended laser system or use the password protected
Screen Lock Button on the Touch Screen/Display Panel to prevent unauthorized use.
Oxygen (with or without other gases)
Never direct oxygen (with or without other gases) toward or over the laser field.
Select the appropriate size mask for patient. Masks with soft or filled cushions help to
minimize leakage.
An oxygen analyzer may be used to check concentrations around the oxygen source
(mask, cannula or airway).
Hair
Whenever treating near the hairline, eyebrows or any other facial or body hair, the hair
must be kept moist with water or saline.
Consider shaving hair-bearing areas (beards, mustaches, arm or leg hair, etc.) prior to
laser treatment to reduce the risk of igniting hair.
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...