Chapter 5 Web-Based Operations
168
Lynx DSA User's Manual - 9240227J
Verify the USB Connection
To verify the USB connection, and to access the Lynx system.
1.
Disconnect your Ethernet cable connection, if it is connected (temporarily for
testing the USB connection).
2.
Connect the USB cable from your PC to the Lynx, and activate your Internet
Browser.
3.
In the URL line, type in
10.0.0.2
, and press
Enter
to connect.
The Lynx’s main access page should open.
RS-232 Connections
Note: This section applies only to legacy 32-bit Windows Operating Systems. This
communication interface is not supported by 64-bit Operating Systems.
A connection for RS-232 communications for either a wired connection or using an
external modem requires Microsoft’s Remote Access Services (RAS) transport
protocol to establish the connection. This means that a browser running on the PC
can be used to access and control the Lynx.
For simplicity in describing the different RS-232 connections in this manual, a wired
RS-232 connection is described as a
Serial Connection
, and an RS-232 connection
using an external Modem is considered a
Modem Connection
.
Setting Up for a Serial Connection
Microsoft’s Remote Access Services (RAS) is used to connect with Lynx via a wired
RS-232 connection. Once configured, a browser running on the PC will be used to
access the Lynx.
The factory-default RAS settings for the Lynx are outlined in the sections below. By
default, the RAS interface is enabled and the factory-default settings can be used to
establish an initial RAS connection between the PC and the Lynx. To change any
parameter from its factory-default settings, the Lynx’s alternate web pages must be
accessed through the direct Ethernet connection using the “crossover” cable
discussed in earlier steps.
The alternate HTML web page in the Lynx are accessed through the following URL:
http://10.0.0.3/main.asp?html5=1