05103-LC AND 05305-LC R.M. YOUNG WIND MONITOR FOR METDATA1

2

4.2  WIND DIRECTION INSTRUCTION CODE

MetData1 uses Instruction 4 (excite, delay, and
measure) to measure wind direction.  In
general, a delay of 2 (0.02 seconds) is sufficient
when lead lengths are less than 100 feet.  If the
cable is 100 to 1000 feet, use a delay of 20
(0.20 seconds).  If you need further assistance
with the delay, contact Campbell Scientific's
Marketing Department.  The wind direction
potentiometer has a five degree deadband
between 355 and 360 degrees; therefore, the
multiplier is 355/Excitation Voltage.  The offset
is zero.

MetData1 Datalogger Program Instruction

01:

P4

(Excite, Delay, Measure)

01:

1

Rep

02:

5

2500 mV slow

03:

3

In Chan

04:

2

Ex Chan

05:

2

Delay (Units 0.01 Sec)

06: 2500

mV Excitation

07:

*

08:

0.142

Mult

09:

0

Offset

* Proper entry will vary dependent on input

location assignment.

5.  MAINTENANCE AND REPAIRS

R.M. Young suggests the anemometer bearings
be inspected at least every 24 months.  Please
refer to the R.M. Young manual for
maintenance information.

Contact R.M. Young's customer service
department directly to obtain repair cost
estimates and authorization for return of the
unit.  Their phone number is (616) 946-3980.

6.  REFERENCES

Gill, G.C., 1973: The Helicoid Anemometer
Atmosphere, II, 145-155.

Baynton, H.W., 1976: Errors in Wind Run
Estimates from Rotational Anemometers  Bul.
Am. Met. Soc., vol. 57, No. 9, 1127-1130.

Содержание Young 05103-LC

Страница 1: ...INSTRUCTION MANUAL 05103 LC and 05305 LC R M Young Wind Monitor for MetData1 4 97 C o p y r i g h t c 1 9 8 4 1 9 9 7 C a m p b e l l S c i e n t i f i c I n c ...

Страница 2: ... in lieu of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization The following contact information is for US and International customers residing in countries served by Campbell Scien...

Страница 3: ...its use inside buildings 2 MOUNTING REQUIREMENTS The Wind Monitor mounts to a vertical piece of 1 IPS schedule 40 1 32 O D pipe See Figure 2 1 A band clamp at the base of the sensor is tightened to secure the sensor to the pipe Campbell Scientific supplies a 12 length of unthreaded pipe for mounting the wind monitor to Campbell Scientific s 019ALU Crossarm Sensor Mount 3 CONNECTION The wind monito...

Страница 4: ...ero MetData1 Datalogger Program Instruction 01 P4 Excite Delay Measure 01 1 Rep 02 5 2500 mV slow 03 3 In Chan 04 2 Ex Chan 05 2 Delay Units 0 01 Sec 06 2500 mV Excitation 07 08 0 142 Mult 09 0 Offset Proper entry will vary dependent on input location assignment 5 MAINTENANCE AND REPAIRS R M Young suggests the anemometer bearings be inspected at least every 24 months Please refer to the R M Young ...

Страница 5: ...This is a blank page ...

Страница 6: ... campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Te...

Отзывы: