9. To query the sensor for its current SDI-12 address, type ?! and press Enter. The sensor
responds with its SDI-12 address. If no characters are typed within 60 seconds, the mode is
exited. In that case, simply type SDI12 again and press Enter.
10. To change the SDI-12 address, type aAb!, where a is the current address from the previous
step and b is the new address (see
(p. 28). Press Enter. The sensor changes its
address and responds with the new address.
11. To exit SDI-12 transparent mode, click Close Terminal.
FIGURE C-2. CR200(X) example of using the SDI-12 transparent mode to change the SDI-
12 address from 0 to 1
WindSonic4 Two-Dimensional Sonic Anemometer
28