PVS5120-Series Portable Samplers
16
FIGURE 6-8. Time Settings tab
If running with time-based control, the user must sync the
clock of the sampler. This is described in Section
(p. 10)
•
Start Date and Time
: Specifies the date and time at which sample
collection begins. Following this point in time, the schedule for all
future samples use the interval specified by the Interval Hour and
Interval Minutes settings. This setting is ignored unless the sample
Control Type
is set to
Time
. The default start date and time are 12:05
AM on January 1, 1990. This must be changed.
If the default Start Date and Time is not changed, the
sampler will immediately begin sampling after the
configuration is applied to the VSC100. Make sure to set
this start date and time far enough into the future to ensure
that the sampler is in place before it starts taking samples.
•
Interval Hour
: Specifies the number of hours in the time interval
between sample collections. This must be an integer value.
•
Interval Minutes
: Specifies the number of minutes in the time
interval between sample collections. Must be an integer value from 0
to 59.
6.2.6 Terminal Tab
The
Terminal
tab allows the user to complete advanced troubleshooting with
the help of a Campbell Scientific application engineer. This tab is no longer
available once the user interacts with the
Settings Editor
tab.
A
NOTE
A
CAUTION
Содержание PVS5120 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 70: ...Appendix A Sample Transport Velocity A 2...
Страница 78: ...Appendix B Example Programs B 8...
Страница 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Страница 96: ...Appendix G Generic Modbus Control G 4...
Страница 100: ...Appendix H ISO5667 Conformity H 4...
Страница 101: ......