Appendix B. Example Programs
B-5
B.4.2 Cycle Shutter/Wiper for Each Measurement CR1000
Program
The following CRBasic program will:
•
Open the shutter if closed, then make a measurement
•
Make a measurement if open, then close
Shutter/wiper cycles will be cut by 50%. This will reduce wear and power
consumption 50% but still leave the optics shuttered 50% of the time.
CRBasic Example B-5. Cycle Shutter/Wiper for Each Measurement CR1000 Program
'CR1000 Series Datalogger
'OBS500 cycle shutter each measurement
'Declare Public Variables
Public
OBS500(4)
Public
obsDatOpen(4),obsDatClose(4)
Public
Open
'Declare Other Variables
Alias
OBS500(1) = turb_bs
Alias
OBS500(2) = turb_ss
Alias
OBS500(3) = tempC_obs500
Alias
OBS500(4) = wet_dry
Alias
obsDatOpen(1) = Open_counts
'Full movement of slider is about 20,000 counts. If it jams this # will be smaller
Alias
obsDatOpen(2) = Open_Max_mA_cnts
' Number of times the shutter stops while opening because of max current
Alias
obsDatOpen(3) = Open_slip
'Open timeout count. If the threads are stripped the slide will not move and this count will increase
Alias
obsDatOpen(4) = Open_mA
'mA current of the motor
Alias
obsDatClose(1) = Close_counts
'Full movement of slider is about 20,000 counts. If it jams this # will be smaller
Alias
obsDatClose(2) = Close_Max_mA_cnts
'Number of times the shutter stops while opening because of max current
Alias
obsDatClose(3) = Close_slip
'Open timeout count. If the threads are stripped the slide will not move and this count will increase
Alias
obsDatClose(4) = Close_mA
'mA current of the motor
Units
turb_bs = fbu
Units
turb_ss = fnu
Units
tempC_obs500 = degC
Units
wet_dry = YesNo
'Define Data Tables
DataTable
(Test,1,1000)
DataInterval
(0,5,Min,10)
Sample
(1,turb_bs,FP2)
Sample
(1,turb_ss,FP2)
EndTable
'Main Program
BeginProg
Scan
(60,Sec,0,0)
'If open make measurement and close. If closed, open then make measurement.
If Open = 1 Then
'If open the make measurement, then close
SDI12Recorder
(OBS500(),1,0,"M4!",1,0)
'Measure without moving the wiper
SDI12Recorder
(obsDatClose(),1,0,"M7!",1,0)
'Close wiper
Open = 0
Else
'if closed
SDI12Recorder
(obsDatOpen(),1,0,"M3!",1,0)
'Open wiper
Delay
(0,11,Sec)
SDI12Recorder
(OBS500(),1,0,"M4!",1,0)
'Measure without moving the wiper
Open = 1
EndIf
'Call Output Tables
CallTable
Test
NextScan
EndProg
Содержание OBS500