This section summarizes some of the factors that affect OBS measurements
and shows how ignoring them can lead to erroneous data. If you are certain that
the characteristics of suspended matter will not change during your survey and
that your OBS was factory-calibrated with sediment from your survey site, you
only need to skim this section to confirm that no problems have been
overlooked.
8.1 Particle Size
The size of suspended sediment particles typically ranges from about 0.2 to
500 μm in surface water (streams, estuaries, and the ocean). With size, shape,
and color remaining constant, particle area normal to a light beam will
determine the intensity of light scattered by a volume of suspended matter.
Results of tests with sediment shown in FIGURE
indicate a wide range of
sensitivity is associated with fine mud and coarse sand (about two orders of
magnitude). The significance of these results is that size variations between the
field and laboratory and within a survey area during monitoring will produce
shifts in apparent TU and SSC values that are unrelated to real changes in
sediment concentration. FIGURE
shows the difference in apparent
turbidity that can result from different ways of disaggregating sediment.
Shutter
copper plate
Floating nut
spacer
Floating
nut
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......