Appendix B. MS-80 example
program
(p. 13) provides wiring for the example program. The MS-80 requires a completion
resistor such as the 3WHB10K to connect to the data logger. Although this program is for the
CR1000X, other CRBasic data loggers are programmed similarly.
Programming tips are provided at:
www.campbellsci.com/videos/datalogger-programming
Table B-1: Wiring for example program
Wire color
Function
CR1000X connection
3WHB10K
connection
Red
Pyranometer signal high
1H
Black
Pyranometer signal reference
1L
Clear
Shield
⏚
(analog ground)
White
PT100 voltage excitation
H
Green
PT100 signal
L
Yellow
PT100 ground
⏚
2H
HI pin
2L
LO pin
⏚
(analog ground)
GND pin
VX1
Black wire
MS-80 Secondary Standard Pyranometer
13