Table 8-2: Analog voltage measurement speed
1
f
N1
(Hz)
Multiplexed
2
measurement
With input reversal
Without input reversal
Time
(ms)
Rate
(Hz)
Time
(ms)
Rate
(Hz)
30000
1.46
698.49
0.75
1394.05
15000
1.53
667.41
0.78
1332.15
7500
1.66
612.87
0.85
1223.49
3750
1.93
526.78
0.98
1051.89
2000
2.40
422.83
1.22
844.59
1000
3.40
297.18
1.72
593.82
500
5.40
186.39
2.72
372.58
100
21.40
46.81
10.72
93.60
60
34.73
28.82
17.38
57.63
50
41.40
24.18
20.72
48.35
30
68.06
14.70
34.05
29.40
25
81.40
12.29
40.72
24.58
15
134.73
7.42
67.38
14.85
10
201.40
4.97
100.72
9.93
5
401.40
2.49
200.72
4.98
2.5
801.40
1.25
400.72
2.50
1
Default settling time of 500 µs.
2
Refers to multiplexing circuitry internal to the VOLT 108/116.
8.2 Resistance measurements
Resistance measurements for four- and six-wire full-bridge and two-, three-, and four-wire half-
bridge using voltage excitation or direct resistance measurements using current excitation.
Excitation polarity reversal minimizes DC error.