o
The transducer faces are fragile. Avoid scratching or rubbing the surface of the
transducer.
o
Grounding the CSAT3BH is critical. Proper grounding to earth will ensure maximum
electrostatic discharge (ESD) and lightning protection and improve measurement
accuracy. The data logger, CSAT3H heater controller enclosure, and CSAT3BH
electronics should all be grounded to the same grounding rod.
o
During winter, remove wicks from the transducers to prevent ice from melting and
collecting on the wicks, which could lead to re-freezing and ice buildup.
o
Avoid opening the heater controller box because of unprotected electronics that are
vulnerable to ESD damage.
l
IMPORTANT
o
Install USB drivers and
Device Configuration Utility
before attaching the sensor to a
computer.
3. Initial inspection
Upon receipt of the CSAT3BH, inspect the packaging and contents for damage. File any damage
claims with the shipping company. Contact Campbell Scientific to facilitate repair or replacement.
Immediately check package contents against shipping documentation. Thoroughly check all
packaging material for product that may be trapped inside. Contact Campbell Scientific about
any discrepancies. Model numbers are found on each product. On cables, the model number is
often found at the connection end of the cable. Check that correct lengths of cables are received.
The CSAT3BH ships with:
l
Temp/RH sensor with radiation shield
l
CSAT3BH replacement desiccant
l
CSAT3BH foam case set
l
CSAT3H heater controller box
l
CM250 leveling mounting kit
l
USB data cable
CSAT3BH Three-Dimensional Heated Sonic Anemometer
2